47
7.4 Moisture trap replacing (furnace)
CAUTION
C9.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
−
Parts of the analyzer can be very hot.
•
Use heat protecting gloves.
1.
Turn both screws (
1
) counter clockwise, until the ring (
3
) touches the
part (
2
).
2.
Lift the glass tube (
5
) upwards, tilt it to the side and pull downwards.
3.
Replace the magnesium perchlorate (Anhydrone). See chapter
4.
Install the glass tube (
5
) in reverse order after greasing the lower O-
ring.
5.
Turn both screws (
1
) clockwise until the O-ring (
4
) is properly pressed
on the inner surface of the glass tube (
5
).
Remark
The O-ring imprint on the whole circle length of the glass tube should be of about 2 mm width.