Data Logger DL210
2
Elster-Instromet GmbH
1 Display
Basic layout of the display:
Archive Device status
M e n u
m a x
á
A W B I n p u t 1
à
Submenu
V 1 A 1 2 3 4 5 6 7 . 8 m 3
Both lines in the display are subdivided into fields which are described below.
1.1 Line 1 = Labels
The first line is subdivided into the following five fields:
1. Type of computation (the first three characters without labels on the front panel)
The type of computation identifies so-called "initial values" (also termed "capture values").
These are values which have been formed over a time period (e.g. the adjustable measure-
ment period or one month). Labels:
F
max
Maximum –
highest value within the time range
F
min
Minimum –
lowest value within the time range
F
D
Change
– volume within the time range
F
Æ
Mean
– mean within the time range
2. Archive
If an arrow points upwards to the label "Archive", then the displayed value is an archived
value. This was frozen at a defined point in time and cannot be changed.
3. Device status
Here a maximum of three of the most important items of status information are continually
shown.
A flashing character signifies that the relevant state is still present and the relevant message is
present in the momentary status.
A non-flashing character signifies that the relevant state is past, but the message in the status
register has not yet been cleared.