905U-E Wireless Ethernet Bridge
User Manual
Man_905U-E Rev 1.0
Page
18
My Right Boot 2.1
Copyright 1999-2004 Cybertec Pty Ltd, All rights reserved.
This software is provided by Cybertec ``as is'' and with NO WARRANTY.
http://www.cybertec.com.au/
ROM : 256KB @ 0xffe00000
RAM : 8192KB @ 0x00000000 (140KB / 0x00023214)
ROM Configuration table ... PASSED.
RAM address pattern check . PASSED.
RAM address bus check ..... PASSED.
Product : E900P
Variant : default-variant
Serial No. : default-serial
Release : default-release
Released date :
Released host :
Build date : Oct 12 2004, 17:27:32
Build host :
Boot Flags : no RAM test, no ROM test, bus timer on, wdog on
static IP, auto-boot, net-boot, reset on
local file, no binary load
Boot delay : 1
Boot Filename : /memory/0xffe40000,0x50000
Boot Address : 169.254.101.93
Boot Netmask : 255.255.255.0
Boot Gateway : 169.254.101.1
Boot Host : 169.254.101.1
Boot Mac 0 : 00:12:af:00:00:04
RTE data store .... no error
Setting bus timer (on) and watchdog (on) ... PASSED
eip: mount point /memory
fec0: connected at 10M Half Duplex.
fec0: local ip = 169.254.101.93, server ip = 169.254.101.1
Press ENTER to abort automatic booting ... 30
d)
Check values for Boot Address, Boot Netmask, and Boot Gateway. These values should
be set to reflect those of the PC you are using to configure the unit. If these are correct
skip to step (h). You may check settings again with the
rct
command. For further help,
type the
help
command.
e)
Set Boot Netmask to the same settings as the computer you have the ethernet cable
connected to. This may be performed with the command:
bnm <Type the netmask>
f)
Set Boot Gateway to the same settings as the computer you have the ethernet cable
connected to. This may be performed with the command:
bgw <Type the gateway IP
address>
g)
Choose an IP address for the 905U-E being upgraded. This IP address must be on the
same network as the computer you have connected the ethernet cable to. This may be
performed with the command:
bip <Type the IP address>
h)
Set boot delay to 1 second with the command
bdelay 1
i)
Switch dip-switch on 905U-E to RUN (OFF) position.
j)
Type the command
reset
. The 905UE will reset and start with the network settings you
have entered.