background image

LIFETIME SA-PB 

 

Suction Device 

 

Product Descriptions and Operating Manual 

 

ELMASLAR A.Ş. 

 

 

 

 

 
 

MODELS: 

 

LIFETIME SA-PB 

 

LIFETIME SA-PB 

 

LIFETIME SA-PB 

 

LIFETIME SA-PB 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 
 

Page 

1

 / 

14 

Summary of Contents for LIFETIME SA-PB 1

Page 1: ...LIFETIME SA PB Suction Device Product Descriptions and Operating Manual ELMASLAR A MODELS LIFETIME SA PB 1 LIFETIME SA PB 2 LIFETIME SA PB 3 LIFETIME SA PB 4 Page 1 14...

Page 2: ...n 8 5 3 Ending Aspiration 9 5 4 Charging LIFETIME SA PB 9 6 Hygienic Preparation 9 6 1 Preparation 9 6 2 Cleansing Disinfection and Sterilization 9 6 3 Connecting Pieces 10 7 Controlling Functions 10...

Page 3: ...1 Overview Front view of LIFETIME SA PB device with reusable collecting container Back view of LIFETIME SA PB device Page 3 14...

Page 4: ...Reusable Collection Container Accessories and spare parts Page 4 14...

Page 5: ...visual warning for 6 seconds 2 Product Descriptions 2 1 Intended Use LIFETIME SA PB is a mobile portable electrically operated medical suction pump used temporarily in adults children and infants To...

Page 6: ...ing with the multi use collection container the device must always stand upright because the ball of the overfill assembly cannot safely close the connection to the suction pump and secretions may ent...

Page 7: ...template on the wall and align it with a water gage 3 Mark the holes for the assembly points 4 Make the holes by using a 3 mm sized drill bit 5 Screw the wall stabilizer by using the screws in the pac...

Page 8: ...ve integrated into suction bag by the saliva the device should always be standing vertically If the fluid is scattered onto the overflow valve it becomes air resistant and the disposable aspiration ba...

Page 9: ...cess ends all the lights on the indicator band will light up steadily Charging with wall stabilizer 1 Connect LIFETIME SA PB device to the wall stabilizer 2 Charging starts automatically a few seconds...

Page 10: ...d is 20 minutes 6 3 Connecting Pieces Following the cleaning disinfection or sterilization procedures connect the pieces as follows 1 Screw the float of the overfill prevention device into the secreti...

Page 11: ...are Correct the incorrect polar connections and reverse change the fuses if required Connection between circle plate and battery Do the interconnection entirely group is not well inserted Device is no...

Page 12: ...ck is full but the charge status of the battery pack will not be displayed unless calibration is performed 12 Do a functional check see the chapter 7 2 Function Controlling 8 1 2 Calibrating capacity...

Page 13: ...vacuum pressure in 12 V 650mmHg in sea level Aspiration hose 7 12 mm Length 1500 mm reusable container Bacteria filter Filter component of 1 m Motor power 55 W Rated voltage 12 V Maximum power 8 A Whi...

Page 14: ...ransport damage caused by improper packaging when the device is returned Not performing the required or stipulated maintenance procedures Normal erosion due to operating The following parts are under...