3.4 KEYS 

 
The front panel is provided with function keys, 

+

 and 

-

 are accelerating when they are kept pressed.  

 

 
 

KEY 

FUNCTION 1 

FUNCTION 2 

M

 

 
 
Steps down the 
menu 

 

 

 
 
Increase numeric 
setting 

 

Interrupt sampling when pressed together with  
 

 

 
 
Decrease numeric 
setting 

 

Interrupt sampling when pressed together with  

 

=

 

 
 
Confirm settings 

 
 
Start sampling 

 

4. SETTINGS 

4.1 SETTING OF SAMPLING CONTROL 

 
Sampling is carried out after setting of the sampling parameters.  

4.1.1 Internal control 

 

A: Manual sampling

 

In manual sampling mode, a sample is taken at the operators command. 

 
1. Press 

+

 and 

-

 at the same time to enter the menu. 

2. Step to 

MAN SAMPLE

 

with 

M

.

  

3.  Start a sampling sequence with 

=

.

 

 
B: Time related sampling 

 

Flow on / Flow off function: 

When time related sampling is chosen there is a possibility to control the sampling with a level control sensor 
(or other potential free closing contact). The function is suitable in conditions where the water level can be low. 
 
FLon: 

Time related sampling is only activated when the level control sensor is activated (see 
connection instructions for the level control sensor) 

FLoF: 

Time related sampling is activated independent of the level control. 
Samples are taken at a pre-set time interval (1 - 36 samples/ hour).  

 
 

M

 

+

 

_

 

=

 

+

 

+

 

_

 

_

 

Summary of Contents for VA/V3

Page 1: ...Water Sampler VA V3 v 2 6...

Page 2: ...ECTION 5 3 3 5 3 3 1 INTERNAL 5 3 3 2 PULSE 5 3 3 3 ANALOG 6 3 3 4 MAX FLOW 6 3 3 5 INTERVAL 6 3 3 6 MAN SAMPLE 6 3 3 7 TOTAL SAMPLES 6 3 3 8 ALARM 6 3 4 KEYS 7 4 SETTINGS 7 4 1 SETTING OF SAMPLING CO...

Page 3: ...e front panel The LCD display gives you running information during both programming mode and sampling sequences 1 1 2 Sampling unit The sampling unit consists of a sample bottle in and outlet tubes an...

Page 4: ...rackets at the rear side of the sampler NOTE Avoid placing the sampler in direct sunlight 2 3 ELECTRICAL INSTALLATION External control For connection to external control a 4 wire cable with Amphenol c...

Page 5: ...2 3 3 Connection of a potential free contact level sensor etc Connection is made through the black and black wire on the external control cable 2 4 CHECKING THE CONNECTIONS Before the main supply is t...

Page 6: ...hen water is present in the vacuum tube the sampler immediately turns off all active functions and empties the sample bottle Error code E 08 is shown on the display In this case contact Elmacron AB 3...

Page 7: ...he operators command 3 3 7 TOTAL SAMPLES Every correct sample without error is accumulated 3 3 8 ALARM When an error appears the red diode at ALARM lights up At serious errors the display flashes and...

Page 8: ...le is taken at the operators command 1 Press and at the same time to enter the menu 2 Step to MAN SAMPLE with M 3 Start a sampling sequence with B Time related sampling Flow on Flow off function When...

Page 9: ...the display flashes 5 Stop the flashing with or Put the decimal point at the correct place with at the left and at the right 6 Confirm with 7 The sampler steps automatically to INTERVAL 8 Set the num...

Page 10: ...ling with 4 2 ALARM EMPTY SAMPLE BOTTLE 1 Press and at the same time to enter the menu 2 Step to ALARM with M 3 Choose between ALon alarm on and ALoF alarm off use key 4 Confirm and start sampling wit...

Page 11: ...on 5 1 SAMPLING SEQUENCE During sampling sequence you can follow the steps in the sampling sequence at the display Display 1 The discharge valve closes 2 Suction tube purge 3 Sample in 4 Discharge of...

Page 12: ...n the screw at the bottle holders upper part 4 Release the bottle pull it straight up 5 If the bottle holder needs to be cleaned loosen the screw at the bottle holders lower part and pull the holder s...

Page 13: ...vacuum tube NOVA if it is OK CH06 Test of menu diodes All diodes on the menu line lights up 6 3 CALIBRATION The water sampler is always pre calibrated when delivered Pre calibration is only necessary...

Page 14: ...l in the tank Check tubes and connections E 03 Sequence collision to many pulses Reduce the number of pulses E 04 The pulse length is to long 5 s Check pulse transmitter E 05 The pulse length is to sh...

Page 15: ...Number of samples 1 36 samples h Main supply 220 V Protection IP65 Max suction height 6 m Alarm output Potential free contact max load 24VDC Control Internal time External potential free pulse Externa...

Reviews: