13
7. TROUBLESHOOTING
Press
=
to accept an error message The error still remains and should be corrected before sampling is starting
again.
CODE INDICATES
CORRECTION
E-01
Dirty sample bottle
Clean the sample bottle
E-02
No water in the tank
Check the water level in the tank
Check tubes and connections
E-03
Sequence collision: to many
pulses
Reduce the number of pulses
E-04
The pulse length is to long> 5 s Check pulse transmitter
E-05
The pulse length is to short < 5
ms
Check pulse transmitter
E-06
Interval > max flow
Decrease the interval
E-07
Max flow / interval > 36 samples
/ h
Increase the interval
E-08
Water in the vacuum tube
Overflow protection is activated
Contact Elmacron AB
E-99
Memory error
Contact Elmacron AB