12 Operating the unit
Elmasonic Select
23 / 32
To switch the heating off, press the
Minus
button until - - flashes on the display. The
Heating
symbol disappears from the display when the heating is switched off.
12.5
Set the ultrasound mode
Degas, eco, sweep, pulse or dyn (dynamic)
Press the
Mode
button to set the ultrasound mode. Press the button repeatedly until
the required ultrasound mode is shown on the display.
12.6
Set auto-degas
Use this function to degas a fresh batch of cleaning fluid.
One of the ultrasound modes
eco
,
sweep
,
pulse
or
dyn
has been set.
The unit is not currently in ultrasound operation.
Hold down the
Mode
button until the ultrasound mode
degas
flashes on the display.
Press the
Start/stop
button to start the
auto-degas
function.
Alternatively, save
auto-degas
as a programme. See the section Save programmes
Degas
flashes slowly on the display (degassing takes approx. 10 min.). Once
auto-
degas
has been completed, the set ultrasonic cleaning mode will start and be shown on
the display.
12.7
Set Temperature-controlled start
Use this function to start ultrasound operation automatically once the target temperature has been
reached.
IMPORTANT
When cleaning sensitive objects, please note that the ultrasound is switched on briefly
multiple times during the heating phase.
The required target temperature has been set.
Ensure that the required target temperature is higher than the current actual temperature. If not,
allow the fluid to cool.
The unit is not currently in ultrasound operation.
Hold down the
Start/stop
button until the
Temperature-controlled start
symbol flashes
on the display.
Press the
Start/stop
button again to start the unit using the temperature-controlled
start function.
See the Save programmes section for information on saving the
Temperature-
controlled start
function as a programme.
The display shows the
Temperature-controlled start
and
Heating
symbols. The display
shows the set target temperature and the current actual temperature of the cleaning
fluid flashing alternately.
During the heating phase, the ultrasound is switched on briefly multiple times to ensure
that the fluid is heated evenly.
The ultrasonic cleaning function will start once the set target temperature has been
reached. The
Temperature-controlled start
symbol will disappear.