When the display shows measurement result are stable, press
H
key to
hold these results.
2. Enter the know values
Press
CF
key to enter the
CF
mode.
Press
SET
key, the display shows the measurement results obtained at
step 1
–
above. If
CF
values are already set, the display shows the
corrected results.
The first line one or more decimal points blinking.
The meter is now ready for you to manually enter the known values.
Press
and
key to select the desired decimal point location. If the set
known value does not have a set decimal point all four of the line
’s decimal
points bling.
Using
and
key, position the cursor on the all parameters element to
adjust.
Using
and
key to change the selected parameter element value.
You can cancel the entry procedure by pressing the
SET
key.
Recheck you entries to be sure that they are right, and then press
key to
store them.
The meter computes the
CF
values by comparing your entered values with the
held measurement values.
If any of the computed
CF
values is lower than 0.0001 or higher than 10.0000,
the meter display shows
“
ErrA
”, and then returns to the beginning of the
SET
mode.
You can view
CF
values by pressing
CF
key to enter
CF
mode and hold down
the
key. The values remain on the screen until you release the key.
If any of the values you have entered are outside of permissible range of the
corresponding parameter, the display shows
“
ErrA
”, and then returns to the
beginning of the
SET
mode. Reenter the values.