Elma 46
24
v. 170309
6.
Press ▼▲key until the desired “Beeper” enable or disable, then press
enter key.
7. Press SET key to exit this mode.
5.4 Calibrating cable length
The cable characteristics parameter is now defined by the factory settings for the cable selected.
But cables are coming from different batches or manufacturers the characteristic variances will
be at up to 20%, to cause deviations in length measurement. To obtain more accurate
measurements, calibrate the tester to the specific cable under test. For proper calibration connect
the cable under test directly to the tester not through a patch cable. To calibrate the tester to the
currently selected cable, perform the following procedure.
1. Select cable type under test.
2. Connect a good cable of known length, >15 m and <100 meters, to the appropriate tester
connector.
3. Press SET key, then press
enter key,
until the display shown “ CAL CABLE ? ”.
4. Press ▼ ▲ key until “YES” is displayed, then press
enter key.
5. Press SET key, “to show cable length, press ▼ ▲ key to adjust the cable length to the exact
length.
6. Press
Key. These cable parameters are stored and will remain in memory even if the test
tool is turned off. All future measurements for this cable type are compared to these new
parameters, until another cable is selected or another calibration is performed.
Hidden calibration cable length mode.
1. Press
key to turn off the meter.
2. Press and hold
down ▼ and ▲ keys then press key to turn on the meter, until LCD
shows
3. Press
▼ ▲ key to cycle display ”
YES
” or ”
NO
”
4. Select
”
NO
” and press ”
” key to enter calibration mode.
5. Otherwise, please select
”
YES
” then press ”
” key to escape.
Reset to factory default
1. Press
key to turn off the meter.
2. Press and hold
”
Test
” & ▼ keys then press key to turn on the meter, until LCD shows
3. Press
▲ key to cycle display ”
YES
” or ”
NO
”
4. Select
”
YES
” and press ”
” (8) key, to enter reset mode.
5. Otherwise, please select
”
NO
” then press ”
”key to escape.
UTP/FTP calibrated parameters is stored in cable category CATX-USER
calibrated parameters is stored in RGX-USER