DC Current
RANGE
Accuracy
Burden Voltage
600.0
µ
A
1)
0.7%+3d
0.25 mV/uA
6000
µ
A
1)
0.5%+3d
60.00mA
1)
0.7%+3d
2.5 mV/mA
600.0mA
1)
0.5%+3d
6.000A
0.7%+3d
0.03V/A
10.00A
2)
0.5%+3d
1)
Ranges for Model 319 only
2)
6A continuous, >6A to 10A for 30 sec. max with 5 minutes cool down interval
Ohm
RANGE
Accuracy
600.0
Ω
,
0.5%+6d
6.000K
Ω
, 60.00K
Ω
0.5%+3d
600.0K
Ω
0.8%+4d
6.000M
Ω
1.0%+5d
60.00M
Ω
1.5%+5d
Open Circuit Voltage: 0.45VDC typical
Audible Continuity Tester
Audible Threshold: Between 10
Ω
and 200
Ω
Response time: 32ms
Diode Tester
RANGE
Accuracy
1.000V
1.0% + 3d
Test Current: 0.50mA typically
Open Circuit Voltage: < 1.6VDC typically
AC Current
RANGE
Accuracy
Burden Voltage
50H
Z
~
500H
Z
600.0
µ
A
1)
2.2%+5d
0.25 mV/uA
6000
µ
A
1)
2.0%+5d
60.00mA
1)
2.2%+5d
2.5 mV/mA
600.0mA
1)
2.0%+5d
6.000A
2.2%+5d
0.03V/A
10.00A
2)
1.2%+5d
1)
Ranges for Model 319 only
2)
6A continuous, >6A to 10A for 30 sec. max with 5 minutes cool down interval