DC

µµµµ

A Current function (Models 079 & 076 only) 

Inputs are made through the test lead terminals COM/+.  

Application notes: 
The  DC

µµµµ

A  function  is  designed  especially  for  HVAC/R  flame  sensor  applications.  The 

0.1

µ

A  resolution  is  useful  for  identifying  the  minute  current  changes  in  flame  detector 

applications.  Flame  signal  current  check  should  indicate  steady  flame  signal  of  at  least 
2

µ

A for a rectification type, or 1.5

µ

A for an ultraviolet type (8

µ

A for self checking systems). 

If  a  flame  signal  current  with  inadequate  strength  or  fluctuation  beyond  10%,  check  the 
following to avoid the risk of unwanted flame relay dropout : 
For gas or oil flames (Minipeeper): 

 

Low supply voltage 

 

Detector location 

 

Defective detector wiring 

 

Dirty viewing windows 

 

Faulty Minipeeper 

For oil flames (Photocell): 

 

Detector location & wiring 

 

Smoky flame or poorly adjusted air shutter 

 

Faulty Photocell 

 

Temperature over 165 

o

f (74 

o

c) at photocell 

For gas flames (Flame Rod): 

 

Ignition interference (A flame signal current difference with the ignition both on and off 
greater than 0.5

µ

A indicates the presence of ignition interference)  

 

Insufficient ground (must be at least 4 times the detector area)  

 

Flame  lifting  off  burner  head  (ground),  or  not  continuously  in  contact  with  the  flame 
rod  

 

Temperature in excess of 600 

o

f (316 

o

c) at the flame electrode insulator causing short 

to ground. 

 

 & 

 3-Phase Rotation function (Models 079 & 076 only) 

Inputs  are  made  through  the  test  lead  terminals  L1/L2/L3.  Phase Rotation  directions  are 

indicated as symbolic movements by the LCD segments. Defaults at 

. Press SELECT 

button momentarily toggles between 

 and 

 modes. 

 :  Hi-sensitivity  mode,  which  detects  relatively  low  signal  outputs  generated  from 

motor spinning, for checking phase connections of Motors.  

 :  Normal-sensitivity  mode  for  identifying  phase  sequence  of  MAINS  (electricity 

supply). 

 

 

Summary of Contents for BM076

Page 1: ...Elma BM076 Dansk Norsk 3 Svenska 16 English 29 EAN 5706445410392...

Page 2: ...ationer 11 Elektriske specifikationer 12 Svensk manual 16 S kerhet 16 Instrumentbeskrivning 17 Anv ndning 18 Underh ll 23 Generella specifikationer 24 Elektriska specifikationer 25 English user manual...

Page 3: ...sikoen for brand og elektrisk st d m instrumentet ikke uds ttes for regn eller fugt Instrumentet er beregnet til indend rs brug For at undg elektrisk st d skal der tages passende forholdsregler ved ar...

Page 4: ...Indikerer hvor str mm ling er mest pr cis 3 Drejeomskifter for t nd sluk og funktionsvalg 4 3 5 6 digits 6000 cifres display 5 Input for fasef lgem ling 6 Input for alle funktioner undtagen m ling me...

Page 5: ...t low pass filter som forbedrer stabiliteten for ACV m ling i st jfyldte milj er For model 073 072 findes EF funktionen p en separat indstilling p drejeomskifteren Detektering af elektrisk felt EF det...

Page 6: ...muligt afstand til motorer transformere og andre ledere under m ling da de kan have indflydelse p m lingen FARE Anvend ikke instrumentet til m ling p str mme h jere end specifikationerne 400Hz K berne...

Page 7: ...lge mellem m lefunktionerne Note N r Diode test funktionen anvendes er det normale sp ndingsfald ved retvendt bias for en velfungerende silicium diode 0 400 til 0 900V Et h jere sp ndingsfald indiker...

Page 8: ...kreds 2 Aflad utlad altid eventuelle kondensatorer f r m ling DC A Str mm ling Model 079 076 B sningerne COM anvendes M ler str m med 0 1 A opl sning 3 faset fasef lge funktion Model 079 076 Input gen...

Page 9: ...s retning er instrumentet tilsluttet faseret I modsat fald skal 2 faser byttes og ny test foretages Akustisk fasef lgesignal Hold knappen REC inde mens instrumentet t nder Herved indikerer instrumente...

Page 10: ...ysningen slukker efter ca et halvt minut for at spare batteri Autosluk Instrumentet slukker automatisk efter ca en halv time uden betjening Tiden resettes af f lgende 1 Drejeomskifter eller tastetryk...

Page 11: ...C Forureningsgrad 2 Opbevaringstemperatur 20 C til 60 C 80 RH uden batterier Anvendelsesh jde Maks 2000m Temperaturkoefficient Nominel 0 15 x specificeret n jagtighed C 0 C 18 C eller 28 C 40 C ellers...

Page 12: ...8 077 Maks 30mm for model 076 073 072 Medf lgende tilbeh r Pr veledningss t manual taske type K tr df ler med bananstik model 079 076 krokodillen bss t model 079 076 Elektriske specifikationer N jagti...

Page 13: ...jagtighed 40 0O C 99 9O C 1 0 0 8O C 100O C 400O C 1 0 1O C 40 0O F 211 8O F 1 0 1 5O F 212O F 752O F 1 0 2O F Type K termokoblers skala og n jagtighed ikke inkl Amptiptm clamp on ACA Skala N jagtigh...

Page 14: ...1 8 5D 100Hz 400Hz 60 00A 4 5 600 0A 2 0 5D 1 Induceret fejl fra n rf rte ledere 0 01A A 2 For model 079 078 077 maksimal Crest faktor 2 1 ved fuld skala 4 1 ved halv skala 3 For model 076 073 072 er...

Page 15: ...ng 3 Tilf j 10D til den specificerede n jagtighed ved 9A Hz frekvens Funktion F lsomhed 1 Sinus RMS Skala 600V 50V 5 00Hz 999 9Hz 60A AMPTIPTM 20A 50 00Hz 400 0Hz 60A 600A 20A 50 00Hz 400 0Hz N jagtig...

Page 16: ...t eller den utrustning man m ter p FARA F r att reducera risken f r brand och elektrisk st t f r instrumentet inte uts ttas f r regn eller fukt Instrumentet r avsett f r anv ndning inomhus F r att und...

Page 17: ...dikator f r k ftarna med DCA polaritetsindikator f r modell 079 078 077 Indikerar var str mm tning r mest exakt 3 Vred f r p av och funktionsval 4 3 5 6 digits 6000 siffrors display 5 Ing ng f r fasf...

Page 18: ...om beskrivs h rnedan utf rs genom anslutningarna COM Standard inst llningen r ACV Tryck p SELECT knappen f r att v lja DCV DC ACV Line level Hz eller EF NCV funktion i n mnd ordning Not ACV och d rmed...

Page 19: ...och fr nkoppling av t ng ppna t ngen och omslut ledaren som skall m tas Var uppm rksam p att k ftarna r helt slutna innan avl sning H ll ledaren centrerad i k ftarna mellan mark rerna f r b sta precis...

Page 20: ...ja mellan m tfunktionerna Not N r diodtestfunktionen anv nds r det normala sp nningsfallet vid r ttv nd bias f r en v lfungerande silikondiod 0 400 till 0 900V Ett h gre sp nningsfall indikerar en def...

Page 21: ...n f r att bl ddra mellan funktionerna Not Vid temperaturm tning skall man alltid vara uppm rksam p att s tta i typ K givaren med korrekt polaritet Varning 1 Anv nd aldrig funktionerna p en sp nningsf...

Page 22: ...i urets riktning f r att generera en signal Om instrumentet indikerar en rotation i urets riktning skall den elektriska matningen L1 L2 och L3 anslutas till motorn s som L1 L2 och L3 indikationerna p...

Page 23: ...rd Kan nollst llningen inte genomf ras pga st rningar uts nds en akustisk signal best ende av 3 korta pip 80ms PEAK RMS Modell 079 078 076 Tryck p PEAK RMS knappen i 1s f r att uppf nga peak RMS str...

Page 24: ...ckage Byte av batteri De 2 skruvarna i batteriluckan p instrumentets baksida lossas ppna batteriluckan och byt batterierna med 2 st nya 1 5V AAA batterier S tt tillbaka och skruva fast batteriluckan G...

Page 25: ...t batteri ca 2 85V f r kapacitans och Hz ca 2 5V f r andra funktioner Autoavst ngning Ca 32 minuter F rbrukn i autoavst l ge Normalt 5 A Dimensioner 223 x 76 x 37mm f r modell 079 078 077 217 x 76 x 3...

Page 26: ...ormalt 1 0V Kapacitans modell 079 078 076 073 Skala Noggrannhet 1 200 0 F 2500 F 2 0 4D 1 noggrannheten r specificerad f r filmkondensatorer eller b ttre Diodtest Skala Noggrannhet 2 000V 1 5 5D Tests...

Page 27: ...inst llning 3 Tillf r 10d till specifikationen vid 4A Standard tang ACA Skala N jagtighed 1 2 3 50Hz 100Hz 60 00A 4 5 600 0A 1 8 5D 100Hz 400Hz 60 00A 4 5 600 0A 2 0 5D 1 Inducerat fel fr n n rliggand...

Page 28: ...00Hz 999 9Hz 60A AMPTIPTM 20A 50 00Hz 400 0Hz 60A 600A 20A 50 00Hz 400 0Hz Noggrannhet 1 5D 1 DC bias inte mer n 50 av sinus RMS Ber ringsfri pols karfunktion Normal sp nning Bargraf indikation 20V To...

Page 29: ...al supply to a building between the building entrance and the main distribution board Such equipment may include electricity tariff meters and primary overcurrent protection devices Terms used in this...

Page 30: ...larity for models 079 078 077 at where best current accuracy is specified 3 rotary selector to turn the power on off and select a function 4 3 5 6 digits 6000 counts numeric lcd display 5 additional i...

Page 31: ...low are made through the test lead terminals COM Defaults at ACV Function Press SELECT button momentarily to select DCV DC ACV Line level Hz or EF NCV function in sequence Note ACV and hence AC DCV Li...

Page 32: ...For non invasive current measurements press the jaw trigger and clamp the jaws around conductor s of only one single pole of a circuit for load current measurement Make sure the jaws are completely c...

Page 33: ...ements Defaults at ACA function Press SELECT button momentarily to select the subject functions in sequence Non Invasive Regular Current Funtions ACA DCA Models 079 078 077 only DC ACA Models 079 078...

Page 34: ...indicates a leaky diode defective A zero reading indicates a shorted diode defective An OL indicates an open diode defective Reverse the test leads connections reverse biased across the diode The dig...

Page 35: ...u can also use a plug adapter Bkb32 Optional purchase with banana pins to type K socket to adapt other type K standard mini plug temperature probes CAUTION 1 Using Capacitance function in a live circu...

Page 36: ...moky flame or poorly adjusted air shutter Faulty Photocell Temperature over 165 o f 74 o c at photocell For gas flames Flame Rod Ignition interference A flame signal current difference with the igniti...

Page 37: ...he phases connected to L1 L2 and L3 of the meter are L1 L2 and L3 also known as R S and T respectively If the meter indicates a counter clockwise movement swap any two connects between the meter and p...

Page 38: ...CD backlight The backlight will also be turned off automatically after 32 seconds to extend battery life Intelligent Auto Power Off APO The Auto Power off APO mode turns the meter off automatically to...

Page 39: ...X 2 Loosen the 2 captive screws from the battery cover case Lift the battery cover case Replace the batteries Replace battery cover case Re fasten the screws General specifications Display 3 5 6 digit...

Page 40: ...w opening 35mm max for models 079 078 077 30mm max for models 076 073 072 Accessories Ttest lead set user s manual soft carrying pouch bkp60 banana plug k type thermocouple models 079 076 only alligat...

Page 41: ...2ms approx Ohm Range Accuracy 600 0 6 000K 60 00K 1 0 5D OPEN CIRCUIT VOLTAGE 1 0VDC TYPICAL Capacitance models 079 078 076 073 only Range Accuracy 1 200 0 F 2500 F 2 0 4D 1 ACCURACIES WITH FILM CAPAC...

Page 42: ...mode applied to offset the non zero residual readings if any 3 Add 10d to the specified accuracy 4A Regular clamp on aca Range ACCURACY 1 2 3 50HZ 100HZ 60 00A 4 5 600 0A 1 8 5D 100HZ 400HZ 60 00A 4 5...

Page 43: ...0A 20A 50 00HZ 400 0HZ ACCURACY 1 5D 1 DC bias if any not more than 50 of Sine RMS Non contact ef detection Typical voltage Bar graph indication 20V TOLERANCE 10V 36V 55V TOLERANCE 23V 83V 110V TOLERA...

Page 44: ...Elma 076 44 MD 20131017 V1 1...

Reviews: