Elma 076
43
Regular clamp-on dc+aca (models 079 & 078 only)
Range
Accuracy
1) 2)
DC, 50HZ ~ 100HZ
60.00A
3)
, 600.0A
2.2% + 7D
100HZ ~ 400HZ
60.00A
3)
, 600.0A
2.7% + 7D
1)
Induced error from adjacent current-carrying conductor: < 0.01A/A
2)
Specified with DC-Zero mode applied to offset the non-zero residual readings, if any
3)
Add 10d to the specified accuracy @ < 9A
Hz line level frequency
Function
Sensitivity
1)
(sine rms)
Range
600V
50V
5.00HZ ~ 999.9HZ
60A (AMPTIP
TM
)
20A
50.00HZ ~ 400.0HZ
60A, 600A
20A
50.00HZ ~ 400.0HZ
ACCURACY: 1%+5D
1)
DC-bias, if any, not more than 50% of Sine RMS
Non-contact ef-detection
Typical voltage
Bar-graph indication
20V (TOLERANCE: 10V ~ 36V)
-
55V (TOLERANCE: 23V ~ 83V)
- -
110V (TOLERANCE: 59V ~ 165V)
- - -
220V (TOLERANCE: 124V ~ 330V)
- - - -
440V (TOLERANCE: 250V ~ 600V)
- - - - -
Indication: bar-graph segments & audible beep tones proportional to the field strength
Detection frequency: 50/60hz
Detection antenna: inside the top side of the stationary jaw
Probe-contact ef-detection: for more precise indication of live wires, such as distinguishing
between live and ground connections, use one single probe to test via terminal com for
direct contact ef-detection with best sensitivity.