BM869s
53
_v61010-2-033:2019
VFD AC Voltage (BM869s only)
RANGE
Accuracy*
10Hz ~ 20Hz
5.0000V, 50.000V,
500.00V, 1000.0V
5.5% + 80d
20Hz ~ 200Hz
5.0000V, 50.000V,
500.00V, 1000.0V
2% + 50d
200Hz ~ 440Hz
5.0000V, 50.000V,
500.00V, 1000.0V
6% + 80d**
*Not specified for fundamental frequency > 440Hz
**Accuracy linearly decreases from 2% + 50d @ 200Hz to 6% + 80d @ 440Hz
dBm
Range and accuracy are subjected to ACmV, ACV, and reference impedance selected.
Typical 600
reference impedance ranges:
At ACmV: -29.83dBm to
–03.80dBm
At ACV: -01.09dBm to 62.22dBm
Input Impedance: 10M
, 60pF nominal
Selectable reference impedance of 4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250,
300, 500, 600, 800, 900, 1000 & 1200
Diode Tester
Range
Accuracy
Test
Current
(Typical)
Open
Circuit
Voltage
2.0000V
1%+1d
0.4mA
< 3.5
VDC
Capacitance
RANGE
Accuracy*
50.00nF
0.8% + 3d
500.0nF
0.8% + 3d
5.000
F
1.5% + 3d
50.00
F
2.5% + 3d
500.0
F**
3.5% + 5d
5.000mF**
5.0% + 5d
25.00mF**
6.5% + 5d
*Accuracies with film capacitor or better
**In manual-ranging mode, measurements not specified below 45.0
F/0.450mF/4.50mF
(450 counts) for
500.0
F/5.000mF/25.00mF ranges respectively