background image

Page 6

ELK-6050 Wireless Smoke Detector

Maintenance

This 6050 Smoke Detector is designed for a service life of approximately
8 to 10 years.  Even so, the following annual and random maintenance
tests are recommended:

Annual

- Smoke Detector Test  - Locate and follow the section in

this manual titled: Full-Alarm Testing of the Smoke Alarm

Random

- Visually check for LED flashing every 16 seconds
- Perform a Detector Sensitivity Test

Detector Sensitivity Test

This test performs a manual diagnostic of the detector with the results
displayed via flashes of the Alarm/Status LED.  To initiate this test:

A. Press and hold the detector Test button for 4 seconds. Once the

test starts, the alarm/status LED will flash 1 to 9 times.

B. Count the LED flashes and view the table below to determine the

results of the sensitivity test and any action required.

IMPORTANT:

If the sensitivity is within limits and all other tests pass, the detector will
go into full alarm mode after the LED flashes stop.  This also causes the
Control to go into alarm.  The detector will auto-reset after 7 seconds,
but the Control's alarm must be manually reset by entering a User Code.
If the sensitivity is not within limits, or an unserviceable hardware fault
is detected, the LED will remain OFF indicating that the Smoke Detector
must be serviced or possibly replaced.

F

FLASHES

MEANING

STEPS TO TAKE

1

Unserviceable hardware 
fault

Reset smoke sensor alarm and rerun 
the sensitivity test.  If error persists, 
replace smoke sensor.

2 - 3

Smoke sensor is becoming 
insensitive

Clear smoke sensor, reset it, and rerun 
the sensitivity test.  If error persists, 
replace smoke sensor.

4 - 7

Smoke sensor is within 
normal sensitivity range.

None.

8 - 9

Smoke sensor is becoming 
too sensitive

Verify smoke chamber is snapped down 
securely.  Clean smoke sensor and 
replace the smoke chamber.

Cleaning the Smoke Detector

Clean the cover with a dry or slightly damp (water ONLY) cloth as needed
to keep it free from dust and dirt. When necessary, use the following
procedure to clean the interior of the detector and replace the optical
chamber (UTC part #211) as follows:
A. Notify the Central Monitoring Station prior to cleaning or testing.
B. Remove the detector from its mounting base.  This will generate a Fire

Trouble condition on the Control annunciated by the keypad.
To temporarily silence the trouble condition press the " * " key on the
keypad.

C. Remove the battery cover and then remove both batteries.
D. Slide a flat-blade screwdriver in the alarm cap slot. Gently push the

handle down to pry the cap up and off.  See figure above.

E. Squeeze the black optical chamber cap at the arrows indicated on

the top of it.  Pull it up and away from the optical base and discard.

F. Carefully blow out all dust and dirt from the optical base.  A soft-

bristled brush may also be utilized.

G. Align the latches on the optical base with the arrows on the NEW

optical chamber and snap it down into place.

H. Replace the detector cap by aligning the four protrusion with the

notches in the detector.  Then insert the cap and turn clockwise about
15 degrees. The cap should snap firmly into place.

I.

Insert the batteries into the detector using the proper polarity, then
replace the battery compartment cover.

J. Reattach the detector to its mounting base.
K. To clear the silenced trouble condition go to the keypad and enter a

valid User Code.

L. Run the Detector Sensitivity test.

Summary of Contents for ELK-6050

Page 1: ...This document should be left with the owner user Owner User Keep this document in a handy location and refer to it when you have questions about this products functions or features Reading this guide...

Page 2: ...you may need to reference other chapters of NFPA 72 or NFPA 101 Inaddition observealllocalandnational building and electrical codes See back page for listing of NFPA Guidelines When choosing an insta...

Page 3: ...personnel identify the sleeping quarters of these individuals IMPORTANT It is recommend that the Smoke Detector be enrolled added into the Control and a Full Test be performed prior to permanently mou...

Page 4: ...r s Test Silence button but down inside the cavity where the heat sensor is located This LED helps provide a way to verify that the wireless RF communications to the Control The LED will briefly flash...

Page 5: ...lowed before a sensor is declared missing 9 PROGRAMTHEZONEDEFINITION ThismustbedonefromKeypad Menu 5 Zone Definitions for all new enrolled wireless zones Program the zone definition of the 6050 detect...

Page 6: ...If error persists replace smoke sensor 4 7 Smoke sensor is within normal sensitivity range None 8 9 Smoke sensor is becoming too sensitive Verify smoke chamber is snapped down securely Clean smoke se...

Page 7: ...total coverage shall include all rooms halls storage areas basements attics lofts spaces above suspended ceilings and other subdivisions and accessible spaces and the inside of all closets elevator sh...

Page 8: ...how two exits from each room It is important that children be instructed carefully because they tend to hide in times of crisis o Establish one meeting place outside the home Insist that everyone meet...

Reviews: