1.1 Instrument Overview
WARNING!
Reagents used in the stainer contain moderately hazardous chemicals that require
care in handling. Always use appropriate safety measures including gloves and eye
protection when handling reagents.
WARNING!
Always wear protective clothing and eye protection when using Nozzle Cleaning
Solution (diluted REF: SS-029C). Dispose of used solution properly.
WARNING!
If power is lost while the stainer is running, the lid will remain locked until power is
restored. Do not attempt to open the lid while power is off.
WARNING!
Electrical shock hazard: Do not open this instrument or attempt internal repairs. Refer
servicing to qualified service personnel. Contact ELITechGroup Biomedical Systems
service.
CAUTION:
This equipment has been designed and tested to CISPR 11 Class A and FCC Part 15 Class
A. In a domestic environment it may cause radio interference, in which case, you may
need to take measures to mitigate the interference.
CAUTION:
To avoid serious instrument damage, always use reagents supplied by ELITechGroup.
Except for approved methanol, using reagents not supplied by ELITechGroup may void
the warranty.
CAUTION:
Only spare parts supplied or specified by ELITechGroup should be used in this device.
Using non-approved parts may affect the performance and safety features of the
instrument. If the equipment is used in a manner not specified by ELITechGroup the
protection provided by the equipment may be impaired. If in doubt, contact your
ELITechGroup representative.
Functional Description
The Aerospray Hematology Stat Slide Stainer/Cytocentrifuge (Model 7122) is a dual-
purpose, microprocessor-controlled slide staining and cell preparation system. In use,
atomizing spray nozzles apply fresh reagents onto microscope slides prepared with
blood or other body fluid specimens. The slides are mounted in a rotating carousel for
processing.