SECTION 4
OPERATING THE STAINER
4.1 Operating Instructions
NOTE:
Access the Reagent Information menu by pressing the bottle icons on the right
side of the Main menu. Press
Change
to scan or manually enter reagent information
.
NOTE
: Do not put residual reagent from a used bottle into a new bottle. This can lead to
an accumulation of residue on the slides and may be a source of contamination.
WARNING!
Reagents used in this instrument contain moderately hazardous chemicals that
require care in handling. Always use appropriate safety measures, including gloves
and eye protection, when handling reagents.
Replacing a Reagent Bottle
1 Remove the empty reagent bottle from the tray but do not disconnect the
dip tube.
2 Open the new bottle and record the letter on the cap for future use, such
as long-term storage.
3 Open the Reagent Information menu by pressing the reagent bottle icon
on the right side of the Main menu.
4 Select the desired reagent and press
Change.
5 If you are using reagent tracking, scan the barcode or manually enter the
reagent REF, expiration date and lot number in the Reagent Information
menu (Section 3.2).
6 Unscrew the cap and remove the dip tube from the empty bottle.
7 Insert the dip tube into the new reagent bottle and screw on the cap.
8 Place the new bottle in the tray.
Emptying the Waste Container
The Level Detect function automatically monitors the waste level and indicates when the
waste container should be emptied. It is still necessary to check waste levels visually to
ensure the waste container does not overfill.
CAUTION:
Dispose of collected waste according to local statutes and safety requirements.
1 Unscrew the cap from the full waste container.
2 Discard the waste according to local regulations.
3 Reinstall the cap on the empty waste container.