ELIS PLZEŇ a. s.
Project design, installation and service manual
Page 26 o
f
33
Electromagnetic flowmeters of the type series
FLONEX FXx11x and FLONET FH30xx
Meter control instructions
ELIS PLZE
Ň a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722
Es 90 686 K
Here the value set by buttons
► and ▲ defines a zone of flow rate symmetric with respect to zero where
all outputs (current, OUT1 and OUT2 outputs, and the magnitude of instantaneous flow rate shown on the
meter display) indicate zero flow rate.
Any flow rate values greater than
Suppressed flow rate
will be correctly measured and indicated.
Return to Basic screen:
ESC 2 s
The suppressed flow rate value shall always be set in litres per second. Conversion of volume flow
rates in various units to flow rate in litres per second is described above in the procedure for setting
the
100%
value.
The initial setting of the suppressed flow rate has been done by the manufacturer according to the
order number specification.
Flow rates for various sensor sizes
DN
NPS
(l/s)
Q
1
v = 0,025 m/s
Q
4
v = 10 m/s
15
½“
0,0036
1,8
20
¾“
0,0067
3,33
25
1
“
0,0100
5
32
1
¼“
0,0167
8,33
40
1
½“
0,0250
12,5
50
2“
0,0400
20
65
2 ½“
0,0667
33,33
80
3“
0,1000
50
100
4“
0,1556
77,77
125
5
0,2389
119,44
150
6“
0,3611
180,55
200
8“
0,6389
319,44
250
10“
1,0000
500
300
12“
1,4000
700
350
14“
1,9444
972
400
16“
2,5000
1250
500
20“
4,0000
2 000
600
24“
5,5556
2 778
700
7,7778
3 889
800
10,0000
5 000
900
12,7778
6 389
1000
15,5556
7 778
1200
22,2222
11 111