2.
SAFETY
2.1
SAFETY INFORMATION
Accessory equipment connected to the analog and digital interfaces must be certificated
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing
equipment and IEC/EN 60601-1 for medical equipment).
Furthermore all configurations shall comply with the system standard IEC 60601-1-2:2014
Everybody who connects additional equipment to the signal input part or signal output part
configures a medical system, and is therefore responsible that the system complies with the
requirements
of
the
system
standard
IEC
60601-1:2005(Thrid
E
CORR.1:2006+CORR.2:2007+A1:2012
If in doubt, consult the technical service department or your local representative.
.
ISO 15004
This report provides information about the hazard to the examinee’s eyed in compliance
with ISO 15004-1:2006, ISO 15004-2:2007 Ophthalmic instruments
– Fundamental
requirements and test methods Part2
– Light hazard protection.
This condition is satisfied
even when the instrument is operating at maximum light intensity and maximum aperture!
(Maximum intensity is the highest brightness the instrument is capable of delivering,
including the highest brightness achievable if overvoltage is provided) detailed radiation
information at normal usage of this instrument is like bellows.
Radiation output: below
117.1 μW/cm2 Limit by ISO15004: 20 mW/cm2
Number
Radiation output [μW/cm2]
1
107.0
2
117.1
3
115.5
4
115.7
5
103.6
6
103.7
7
108.8
8
109.0
9
105.6
10
105.8
average
109.1
.