6
Operating the Device
104
switch must be ON. If switched to OFF, a message is displayed that the initialization cannot be
performed.
6.9.4 Device Test
After selecting this menu item, the device tests its internal state and displays a list of detected
errors and warnings. The test of the device runs for several seconds and does not affect the
measuring and archiving operation of the device. The execution of the command is independent
of the position of the service switch.
The notification is displayed during the test.
6.9.5 Device Errors Display
Error messages are displayed in the "Actual Status", "Summary Status" and "Device Test"
menus. Autodiagnostics runs regularly, a full test of the device once a day, a sensor swap test
every hour or irregularly when the device is switched on. The test can be also triggered by
selecting the "Device Test" function from the keypad.
The shortened form of summary diagnostics is displayed in the right corner of the top level menu
in the form of OK, Err or Wrn (see Section 6.2). This shortened form is the sum of the individual
states, and in each case, the abbreviation with the highest priority is displayed. The priority order
is Err, Wrn, OK, starting with the highest one. More detailed diagnostic information can be
displayed using the utility SW [42], [43]. An overview of the indicated Err messages is included in
, and the Wrn report is provided in
Displayed Message
bit
Description
E0 program CRC
0
Firmware checksum error.
E1 loader CRC
1
Checker loader error.
E2 parameter CRC
2
Device parameter checksum error.
E3 memory error
3
Device memory error.
E4
- not used -
E5 setting archive full
5
The number of changes made to the device configuration
has filled the archive capacity. The device is still working,
only the parameter changing is blocked. It can be
unblocked in ASC only.
E6 transd. replacement
6
Unauthorized replacement of the sensor or modification of
its parameters has been made.
E7 trans. communication
7
Error in communication with the sensor.
E8 transd. error
8
Transducer error.
E9 battery discharged
9
The EVC battery is discharged (the remaining battery
capacity is about 10%).
E10 compress. table
10
Error in compressibility table calculation due to input
parameters.
E11 compressibility error
17
The compressibility cannot be calculated due to the
limitation of the scope of the standard used for the
compressibility calculating in the measured gas
temperature and pressure.
Summary of Contents for DATCOM
Page 38: ...3 Technical Parameters 36 ...
Page 50: ...5 Installation and Commissioning 48 Fig 7 Examples of data plates ...
Page 83: ...5 Installation and Commissioning 81 Fig 37 Communication via RS232 RS485 using B RS module ...
Page 115: ...7 Technical Description of the Device 113 Fig 57 Functional diagram of ELCOR ...
Page 116: ...7 Technical Description of the Device 114 Fig 58 Functional diagram of ELCORplus ...
Page 122: ...7 Technical Description of the Device 120 Fig 61 Modem replacement ...
Page 159: ...10 Communication with the Device 157 packet ...
Page 174: ...12 Accessories 172 Fig 78 ELCORplus example of use of external modules ...
Page 213: ...Appendix 3 Compass Data Export 211 Preview of generated document ...
Page 215: ...Appendix 3 Compass Data Export 213 3 Export to XLS 4 Export to TXT DOC ...
Page 216: ...Appendix 3 Compass Data Export 214 ...
Page 217: ...Appendix 3 Compass Data Export 215 5 Export of graph ...