8.2 Liquid sanitization
The unit is sanitized to reduce the growth of microbiological
contamination within the unit. Please read this entire section to
become familiar with the procedure before you start. Sanitization may
be required in the following circumstances:
To maintain low bacterial counts
If the unit has not been used for a prolonged period of time
If the product water is used for particularly stringent
applications
Step 1
– Start Sanitization Cycle
1. ENSURE unit is ready to use.
2. SWITCH OFF electrical supply.
3. REMOVE POU filter, if fitted.
4. OPEN the door.
Step 3
– Add Minncare Cold Sterilant into LA698-US Liquid
Sanitization Block
1. UNSCREW
the
container
from
the
LA698-US
sanitization block.
2. ADD 10ml of Minncare Cold Sterilant into the container.
There is a line on the container to assist you.
3. REASSEMBLE the LA698-US by screwing the
container back onto the main body.
CAUTION!
ENSURE THE CONTAINER REMAINS
UPRIGHT SO THAT THE CHEMICAL
REMAINS WITHIN THE LA698-US.
4. REFIT the container on to by-pass block, hand tight.