PURELAB Classic
Operator Manual
ELGA
Page 24
PURELAB Classic
Version 2 11/15
Step 5 - Fit Purification Pack
1. REMOVE a new Purification Pack from its packaging.
2. REMOVE the sealing plugs from inlet and outlet ports.
3.
WET ‘O’ rings and SLIDE new Purification Pack into the
right hand position pushing upwards against pack
reader contacts.
4. POSITION Purification Pack onto spigots.
5. PUSH into unit.
6. ENSURE that the Purification Pack is fully engaged and
dropped down past the retainers.
Step 6 - Acceptance of Purification Pack Installation
1. SWITCH ON power.
2. PRESS TICK
button to accept the Purification Pack.
3. PRESS TICK
button to calculate replacement date.
Step 7 - Final Rinse/Air Bleed
1. PRESS PROCESS
button and ALLOW the process
screen to appear.
2. PRESS Dispense Controller. DIRECT dispense flow to
drain or collect in suitable container (±10 litres).
3. ROTATE Dispense Controller and INCREASE the
dispense rate to maximum.
Note:
The unit will start and initially during this stage
air/water will be purged from the unit via the
dispenser and the drain line.
4. ALLOW to dispense for 4 minutes.
5. RELEASE air from UF (UV) by carefully opening the
valve located near the top of the UF if fitted.
6. PRESS Dispense Controller to stop dispense.
Purification Pack Installation Screens
Retainer
Spigots
Fit Purification Pack