Page 7
Replacing
Removing
5.2 Purification Cartridge (LC207)
1.
SCREEN prompt to install the purification
cartridge (LC207)
2.
REMOVE the LC207 purification
cartridge from packaging.
3.
REMOVE sealing plugs from inlet and
outlet ports
4.
WET ‘O’ rings and SLIDE into right hand
position pushing upwards against metal
pack reader contacts.
5.
ENSURE that the pack is fully engaged
and drops down past the retainers.
6.
PRESS
✓ to continue
.
5.3 Conditioning Cartridge LC206
1.
SCREEN promt to install the conditioning
Cartridge (LC206).
2.
Remove the LC206 conditioning
cartridge from packaging.
3.
REMOVE sealing plugs from inlet and
outlet ports
4.
WET ‘O’ rings and SLIDE into right hand
position pushing upwards against metal
pack reader contacts.
5.
ENSURE that the pack is fully engaged
and drops down past the retainers.
PRESS
✓ to continue and acknowledge
replacement dates for the following:
LC207, CVF (LC136), UV (LC105), PT
(LC177), UMF (LC109), CC (LC206),
Sanitization reminder
.
5.4 System purge
6.
PRESS process button
.
The system now enters an automatic
purge phase.
After a maximum of 60 minutes the
system will exit the purge phase
regardless of the final water purity.