ELGA MEDICA Pro EDI Quick Start Manual Download Page 3

ELGA 

MEDICA Pro EDI - US Quick Start Manual 

 

MEDICA Pro EDI - US MANU40683 Ver.2 

– 10/15

 

 

Page 1

 

1. 

INTRODUCTION 

Use of this manual 

ELGA LabWater strives to produce manuals that are as simple and accurate as possible. However, should you feel 
they can be improved in any way please email us at 

[email protected]

  

This  shortform  manual  guides  you  through  the  installation,  commissioning  and  basic  operation  of  the 

MEDICA  Pro 

EDI

 

–  US 

allowing  you  to  obtain  a  guaranteed  supply  of  purified  water  to  meet  your  requirements.  For  more 

information that is detailed refer to the Operator Manual supplied on CD. 

Customer Support 

Service  support  and  consumable  items  are  available  from  your  local  supplier  or  distributor.  Please  contact 

[email protected]

 for further information. 

 

 

2. 

HEALTH AND SAFETY NOTES 

 

WARNING!

 

WARNINGS  ARE  GIVEN  WHERE  FAILING  TO  OBSERVE  THE  INSTRUCTIONS 
COULD RESULT IN INJURY OR FATALITY. 

 

CAUTION!

 

Cautions  are  given  where  failure  to  observe  the  instructions  could  result  in 
damage to the equipment, associated equipment and processes. 

2.1  Electricity 

It  is  essential  that  the  electrical  supply  to  the 

MEDICA  Pro  EDI

 

–  US

  is  isolated  before  any  items  are  changed  or 

maintenance work performed. 

The appliance coupler (mains lead) connected to the rear of the unit can  be removed to isolate the power supply. If 
access to this is restricted then it is recommended that access to the supply socket is easily available to disconnect 
the electrical supply. 

The unit must be earth. Use of the recommended appliance coupler will ensure adequate earth protection. 

 WARNING!

 

IF  THE  EQUIPMENT  IS  USED  IN  A  MANNER  NOT  SPECIFIED  BY  ELGA,  THE 
PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMPAIRED.. 

 

2.2  Pressure 

The water supply should be isolated and residual pressure released prior to removal of any components or carrying 
out work on the unit. 

Switch off the process and relieve pressure in the distribution loop by opening a point of use or dispense tap. 

2.3  Sanitization Chemicals 

During the automatic sanitization cycles, EfferSan

™ multi-purpose disinfecting tablet and Minncare cold sterilant are 

used  and  the  relevant  safety  information  is  included  within  the  operator  manual  included  on  CD  in  the  rear  of  this 
manual. 

Please refer to the manufacturer for material safety data sheets.  

 

Summary of Contents for MEDICA Pro EDI

Page 1: ...MEDICA Pro EDI US Quick Start Manual...

Page 2: ......

Page 3: ...sses 2 1 Electricity It is essential that the electrical supply to the MEDICA Pro EDI US is isolated before any items are changed or maintenance work performed The appliance coupler mains lead connect...

Page 4: ...d that the weight of other equipment is also taken into consideration Water supply Local potable supply with a means of isolation 3 2 External services and connections UNIT OF MEASURE Connection Termi...

Page 5: ...to enable access to the sanitization port Potable feed water tube connection 2 CONNECT both drains to local drain with tube supplied Note Ensure air gap is maintained from outlet of tube into a standp...

Page 6: ...EDI US Quick Start Manual Page 4 MEDICA Pro EDI MANU40683 Ver 2 10 15 4 CONNECT electrical supply lead into the socket on the rear of the MEDICA Pro EDI unit and to local isolated supply Electrical su...

Page 7: ...ashing CLOCK clock icon 2 PRESS SCROLL to edit time OR PRESS TICK to proceed to step 4 3 3 PRESS and UP to increase or DOWN to decrease hour 4 PRESS SHIFT to step cursor onto minute 5 PRESS UP to incr...

Page 8: ...ocated on the inlet and outlet of the Protek 5 OPEN the left hand door and locate the pre treatment position 6 LIFT wing and rotate anti clockwise 7 PIVOT latch and release retaining arm 8 SWING retai...

Page 9: ...Conditioning Cartridge LC206 1 SCREEN promt to install the conditioning Cartridge LC206 2 Remove the LC206 conditioning cartridge from packaging 3 REMOVE sealing plugs from inlet and outlet ports 4 W...

Page 10: ...lgalabwater com For the address of the nearest ELGA LabWater Sales and Service office visit the country list on our website http www elgalabwater com or contact ELGA at the number above 2015 ELGA LabW...

Reviews: