CENTRA-R 60/120 US Operator Manual
ELGA
Page 22
CENTRA-R 60/120 US MANU40596 Version 02 10/16
5.3
Replacing Labpure L1 purification cartridge
(LC187)
When the purity of water from the unit starts to
deteriorate.
When indicated by the consumable alarm or after
a maximum of six months.
1. ENSURE process is OFF and ISOLATE power.
2. OPEN front doors and LOCATE Labpure L1
cartridge.
3. PUSH Labpure L1 FORWARD (1), then LIFT (2),
finally PULL (3) to remove and discard the
cartridge.
4. REMOVE new Labpure L1 from packaging.
5. REMOVE sealing plugs from inlet / outlet ports
and WET O rings.
6. SLIDE cartridge into position (1) pushing upwards
against the contacts.
7. EASE back and ENSURE that the cartridge is
fully engaged (down) (2) in the retainers.
8. SWITCH on power. FOLLOW screen prompts to
accept new cartridge replacement date.
9. START unit and allow to circulate until water quality
is achieved.