CENTRA-R 60/120 US Operator Manual
ELGA
Page 8
CENTRA-R 60/120 US MANU40596 Version 02 10/16
3.2 System Preferences
When the
CENTRA-R 60/120 US
unit is started for the first
time after completing the commissioning routine the
following steps should be carried out to set up your
choices:
Note:
Your choices can be changed and implemented during
normal operation. It is not necessary to stop the unit.
Step 1 - System Access (User Key)
The User Key prevents unauthorized access to specific
settings. This ensures consistent system performance and
operation.
Note:
The User Key does not prevent access to the
PROCESS function in case of emergency.
1.
SWITCH on the main electrical supply to start the
controller set-up sequence. This takes several
seconds.
2.
PRESENT the Master Key (black) or the User Key
(blue) to the reader ensuring clean contact of both
metallic components.
3.
The display will show a padlock
.
4.
REMOVE Key from the reader.
5.
The display will show a key
.
6.
PRESS
button.
Note:
If no buttons are pressed the system will relock
after 5 seconds.
Step 2 - Clock (screen 018)
1.
PRESS
to edit time
OR
PRESS
✔
to proceed to Step 3.
2.
PRESS and HOLD
to increase or
to
decrease hour.
3.
PRESS
to shift cursor onto minute.
4.
PRESS
to increase or
to decrease minutes.
5.
PRESS
.
6.
PRESS
✔
.
Setting clock time