EN
22/08/22
User Manual Resistance Meter 30A/2
–
60A/2
–
80A/2 62000640_08_
16
2.3.3.
Output Connector Description
RHM DESCRIPTION OUTPUT CANNON 25 POLES FEMALE
1
- common measure ohm/volt
2
- n.c.
3
- n.c.
4
- n.c.
5
- n.c.
6
- n.c.
7
- n.c.
8
- n.c.
9
- command range 2Kohm output 15Vdc
10
- command range 20-ohm output 15Vdc
11
- n.c.
12
- analog GND
13
- +15Vdc
14
- command measure
Ω
(Short circuit with pint 1 to measure
Ω
)
15
- n.c.
16
- n.c.
17
- n.c.
18
- n.c.
19
- n.c.
20
- n.c.
21
- n.c.
22
- command range 200-ohm output 15Vdc
23
- command range 2-ohm output 15Vdc
24
- ohm/volt analog output 5Vdc f.s.
25
- digital GND (reference command range output)
N.B. : to enable the measure of the resistance short circuit the pin1 with the pin14 with
an external contact (that support 15Vdc 50mA). With the open contact has enable
the measure of the Volts.