02/02/22
User Manual CPS/M 62000584_05
20
2.7
MODE MENU
2.7.1
Voltage reaction
The stabilization of the output voltage coincides both on the CPS/M output terminals (2-wires)
and on a possible long distance outlet (4-wires) to eliminate the fall in voltage influence of the
connections.
To operate the long-distance stabilization first connect the SENSE terminals on the back of the
machine following the indications at the voice INSTALLATION.
The choice of the kind of long-distance stabilization can be operated by means of the buttons 4
wires and 2 wires at the voice SENSE.
Notice that CPS/M corrects falls in voltage on the connections until 5% of the set-up voltage to
avoid possible overheating of the line; after exceeding this limit CPS/M
doesn’t guarantee that the
value of the output voltage is equal to the set-up voltage and an error signal appears (see at
voltage alarms).
2.7.2
Output Relay
If you press the button OUT you can turn-on or turn-off the output relay. Before the turn-off the
output voltage is set to 0 voltage so the relay switch with low current. Also, the turn-on is at 0
voltage and after that, the voltage goes to the setting value with an internal ramp.
The turn off of the output relay does not allow operations on the connection of the EUT.
2.7.3
DC/AC mode
By pressing the Mode button, the CPS/M switch from AC to DC and vice versa.
Summary of Contents for 99112103
Page 29: ...02 02 22 User Manual CPS M 62000584_05 29 3 4 WIRING DIAGRAM 3 4 1 2 WIRE CONFIGURATION ...
Page 30: ...02 02 22 User Manual CPS M 62000584_05 30 3 4 2 4 WIRE CONFIGURATION ...
Page 63: ...02 02 22 User Manual CPS M 62000584_05 28 4 5 SCHEMI DI CABLAGGIO 4 5 1 CONFIGURAZIONE 2 WIRE ...
Page 64: ...02 02 22 User Manual CPS M 62000584_05 29 4 5 2 CONFIGURAZIONE 4 WIRE ...