elepho eClear Full User Manual Download Page 9

9

Put the USB plug connected to DC 5V power supply 
into the USB socket at the bottom of the device.

After power is turned on, the unit is powered by the internal lithium battery without the use of the 
USB cable. When fully charged, it can be used for approximately 100 times for one minute each 
usage. The Lithium battery can charge and discharge for approximately 500 times during the life 
of the rechargeable battery.

7.6.1 - Start up

a.  In standby mode, press the power button to turn 

the eClear on at which point all symbols will be 
briefly displayed on the screen. By default level 
one operation will be selected.

b.  Press the button (once) to move to different (higher) 

suction strengths (between I, II and III).

c.  To turn off the device keep the button pressed and 

the device will automatically turn off.

d.  In the charging mode by pressing the power 

button the device will automatically cut off the 
charging mode and switch to the working mode of 
(first suction gear) operation. 
 
The battery power is not displayed at this time. 
 
After use, the nasal aspirator will return to the 
charging mode if connected to the charger.

Notes:

1.  Please choose an adapter that meets the IEC 

60601-1-2 standard.

2.  When the lithium battery power is low, you can 

connect the USB cable with DC 5V input and 
continue to use the nasal aspirator. While in use, 
the nasal aspirator will disconnect the charging 
of the battery. When the nasal aspirator is turned 
off it will switch back to the charging mode if still 
connected to the charger.

7.4 Using the Optional USB power

7.5 Using the internal battery

7.6 Operating instructions

USB

Summary of Contents for eClear

Page 1: ...Full User Manual...

Page 2: ...structions Installation parts description How to install Notice before use Using the optional USB power Using the internal battery Operating instructions Troubleshooting Care and maintenance Warranty...

Page 3: ...ny way that the device will not be affected by electromagnetic interference Avoid using the device in high electromagnetic environment The product is classified as Class II ME equipment when it works...

Page 4: ...this product in the ambient temperature range of 4 F 131 F 20 C 55 C When the equipment is transferred from the lowest storage ambient temperature to the ambient temperature of 68 F 20 C it takes 40 m...

Page 5: ...outh open do not hold your breath or swallow saliva so as to avoid unbalanced ear pressure It is strictly forbidden to clean or soak the body parts under the nasal aspirator collection container Pleas...

Page 6: ...Relative humidity RH 93 Atmospheric pressure 700 hPa 1060 hPa Product weight 5 9oz 167g Suction range 65kpa Automatic Power off 65 s Internal power supply Lithium Polymer Battery DC 3 7V USB Power Sou...

Page 7: ...please check whether the product appearance is good and complete If there is any defect please contact the supplier or manufacturer 7 2 2 Component installation a Align the hole position of silica ge...

Page 8: ...saline sesame oil for about a few minutes and then use the aspirator to suck out the mucus Do NOT use excessive force or hard pulling while extracting a booger from the baby s dry nose b Cleaning and...

Page 9: ...d III c To turn off the device keep the button pressed and the device will automatically turn off d In the charging mode by pressing the power button the device will automatically cut off the charging...

Page 10: ...wall of the collecting cup it is recommended to clean the collecting cup and the transparent cover in advance If the liquid enters the side wall gap and triggers the electrode to enter the alarm mode...

Page 11: ...y dry it and store it in a dry and clean place Note Please make sure that the transparent cover collecting cup and silica gel cover have been dried without any residual water to avoid the drops trigge...

Page 12: ...131 F 20 C 55 C 2 Relative humidity 10 93 Problem Possible Cause Proposed Solution No display when you press the power button The battery may have been depleted if not used for a long time Please use...

Page 13: ...time of purchase Note if the product is disassembled the warranty is void Please keep the sales receipt as evidence of when where the device was purchased Manufactured for Elepho Inc by Hetaida Techn...

Page 14: ...11 Class B Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and Manufacturer s Declaration Electromagnetic Immunity Immunity Test IEC 606...

Page 15: ...e due to repair or tampering by the customer or unauthorized third parties Damage incurred during transport from the manufacturer to the consumer or during transport to the service center Accessories...

Page 16: ...Inc 466 Central Ave 2nd floor Cedarhurst New York 11516 USA Item Device HTD2601US Product model name eClear E mail support customercare elepho com Telephone support via Elepho Consumer Relations at 1...

Reviews: