Error messages
11. Error Messages
After troubleshooting, all error messages can be quit by pressing ESC. If you press Menu/Enter you
will go back to the main menu.
Error Message
Cause
Remedy
Probe or GND error
Probe connecting or earthing error
Check probe cable for damages.
Connect probe cable properly.
Earth equipment properly.
Probe connection error
Data communication failure. Connection
between probe and control unit has been
interrupted during operation.
Check probe cable for damages.
Connect probe cable properly.
Probe voltage drop
Voltage drop between control unit and probe
Recharge battery or switch to mains supply or
attach a shorter connecting cable
Low battery. Please
recharge!
Low battery
Recharge will be necessary soon.
Battery run down. Please
recharge.
Insufficient battery charge
Immediate recharge necessary.
Long bleep - instrument
switches off automatically
Deep discharge protection
Immediate battery discharge necessary.
!
Data have been changed or deleted from
memory
Please send the instrument to your dealer for
repair.