Important Note
In ultrasonic thickness measurement it is inherent to the measuring procedure that the gauge might
use the second echo rather than the first one reflected from the material being measured. This may
result in a thickness reading TWICE as high than the actual thickness.
When measuring through extremely thick coatings in the echo-echo mode, it may occur that the
thickness of coating layer is measured instead of the wall thickness that actually was intended to
be measured.
The responsibility for proper use of the gauge and recognition of this phenomenon rests solely with
the user of the gauge.