3 .2 .3
F ov ea m easu rem en t
In a ZATA test it is possible to measure the sensitivity of the eye at the fovea. For this
measurement the patient fixates the centre of a four-point pattern of lights positioned below
the normal central fixation light.
The software uses a 4-2 staircase algorithm to obtain three measurements of the foveal
threshold, and then takes the average of these three readings to give the final estimate. The 3
measurements are intertwined to avoid any obvious sequences.
1.
Click
and then confirm in the pop up window.
2.
Instruct the patient to look at the centre of the four lights, and to press the response
button when they see a light flash at the centre of the pattern. the operator should
emphasise that many of the presentations will not be seen. If the patient is not sure
they saw a stimulus presentation they should not press the response button.
3.
Click
to start testing.