background image

VD3LQRQ_ING.doc 03/09/2020                                                                                                                                                        2/2 

 

Gain 

The instrument is normally supplied with gain value 

x1; setting gain x2 or x4 allows to reach the FRS with 

half or one fourth of the max potentiometer 

excursion. 

 

Connecting the transducer with one end to zero 

With this configuration the instrument displays zero 

when the cursor is next to one of the transducer 

ends. The ends are linked to terminals 4 and 7 while 

the cursor is connected to terminal 8. 

 

Connecting the transducer with mid-stroke zero 

With this configuration the instrument displays zero 

when the cursor is at the transducer mid-stroke. The 

transducer ends are linked to terminals 4 and 6 while 

the cursor is connected to terminal 8. 

 

Calibration

 

We advise to adjust the readout in the following way: 

  Set the machine axis to zero and adjust the display 

indication  to  zero  by  means  of  the  trimmer  P1 

placed on the front case 

  Set  the  machine  axis  to  its  maximum  excursion 

and  adjust  the  display  at  the  requested  value  by 

means  of  the  display  range  trimmer  P2  placed  on 

the rear case.  

Repeat the operation to achieve the best measuring 

accuracy. 

The  display  value  increasing  direction  can  be 

inverted by inverting the transducer ends. 

REMARK

The  readout  type 

TF

  has  both  trimmers  placed  on 

the  front  case, which    must  be  removed to calibrate 

the  instrument.  To  remove  the  front  case  slightly 

lever  between  the  front  case  and  the  frame  with  a 

small flat screw-driver. 

 

MAINTENANCE

 

Check the  terminals clamping regularly. 

 

 

 

Reviews: