Page 8 of 16
Release 1.0
User Manual
PAT-Chamber-16
4 Installation
Place the PAT-Chamber-16 on a flat, horizontal surface. Do not place the instrument on a
flammable surface. A 230 V or 115 V power connection must be available. The distance
between the wall and the rear of the instrument must be at least 15 cm. The top clearance
must not be less than 25 cm. Sufficient air circulation in the vicinity of the instrument must be
guaranteed at all times.
4.1
Connection to the external battery tester or potentiostat
Two versions of the PAT-Chamber-16 are available which differ in the way the connection to
the external battery tester or potentiostat is being established.
1. PAT-Chamber-16:
Direct connection with fixed
assignment of WE, CE and RE for a
given channel.
2. PAT-Chamber-16 C:
Indirect connection via
PAT-Connect-16 C with variable
assignment of WE, CE and RE for a
given channel.