7.7 Inside the glove box:
Attach the
counter piston from below. Use the
removal tool to push it. The piston must
be pressed into the frit flange until the
two sealing rings are no longer visible.
Make sure that it is positioned straight in
the frit flange.
7.8 Inside the glove box:
Tighten the
socket screw at the end of the piston
firmly using the tools provided.
Important note:
Only use a frit flange with an attached
metal support ring as shown in the
picture. Otherwise the frit flange may
be damaged when trying to tighten the
screw!
7.9 Inside the glove box:
: Put this
assembly into the cell base body. Make
sure that the two grooves at the frit flange
and the cell base body are properly
aligned.
Note:
Make sure that the frit flange is fully
inserted. You may otherwise damage the
cell.