User Manual ECC-Opto-Std – Release 2.62
Pull the syringe piston back to evacuate the cell. Hold the piston a few
seconds in the strained position.
Make sure that the tip of the syringe points downwards. Then release the
piston so that the electrolyte is sucked into the cell by the previously applied
vacuum. NOTE: Never pressurize the cell during the filling procedure.
Repeat the two previous steps to complete filling.
Remove the filling tube and close port B with the second WE contact pin.
11.
Insert the cell into the stand.
12.
Connect the cell to the cell cable as shown in the picture on the following page.