Chapter 7
CAN
B
ROWSER
C
ONTROLS
Below tables 5 and 6 present list of registers that can be accessed and/or read-out using CAN
Browser.
Table 5.
Basic controls
Setting
Description
Command
Toggle the HV supply output ON/OFF
Uset
Set HV output value.
Controller setup/Analog
control
Off – control HV supply via CAN/RS232.
On – control HV supply by potentiometer.
Table 6.
Advanced control registers
Setting
Description
External off status
Inactive – pin
1
of connector CAN/RS232 (
#7
) is
not
connected
to ground.
Active – pin
1
of connector CAN/RS232 (
#7
) is connected to
ground. HV power supply output is turned off.
Udisp
Display HV output voltage.
IsensP
Display positive output current of HV supply.
IsensM
Display negative output current of HV supply.
Tpcb
Display PCB temperature of HV supply.
Vpwr
Display input +48 voltage.
Vprog
Display maximal HV value of output for potentiometer mode.
Interlock/Vpwr max
Display actual value of +48V maximal input. If DC input is over
the maximal value (see
Table 3
), HV power supply turns off.
Interlock/Uout max
Display setting of maximal HV output value. If this value is
exceeded, HV power supply is off.
Interlock/IsenseP max
Display setting of positive output maximal current value. If this
value is exceeded, HV power supply is off.
Interlock/IsenseM max
Display setting of negative output maximal current value. If this
value is exceeded, HV power supply is off.
Interlock/Tpcb max
Display setting of PCB maximal temperature. If this value is
exceeded, HV power supply is off.
Interlock/Tunildm max
Display setting of processor maximal temperature. If this value
is exceeded, HV power supply is off.
Controller setup/Uset limit
Display setting of HV output limit.