background image

 

5

 

www.eksmaoptics.com 

 

 

Figure 3.

 Rear controls of the HV power supply 

 

Table 4.

 Controls and connections of the HV power supply 

Port 

Potentiometer for HV tuning in potentiometer mode 

LED HV output on 

LED +48V 

+24V auxiliary output connector 

5.1 

2 pins for +48V input 

5.2 

2 pins for GND input (-48V) 

6.1 

Test point for -HV current monitoring (1V = 100mA) 

6.2 

Test point for +HV current monitoring (1V = 100mA) 

6.3 

Test point for +HV output monitoring (1V = 1kV) 

6.4 

Test point GND 

CAN/RS232 connector 

Grounding screw 

+HV output 

10 

GND output 

11 

-HV output if equipped.  

Summary of Contents for HV-200

Page 1: ...HIGH VOLTAGE POWER SUPPLIES HV 200 HV 400 HV2 200 Technical Description Rev 2202 2022 Lithuania ...

Page 2: ......

Page 3: ...Models 2 2 1 2 Main Components 2 2 2 TECHNICAL SPECIFICATIONS 3 CHAPTER 3 DEVICE LAYOUT 4 CHAPTER 4 SAFETY 7 CHAPTER 5 QUICK START GUIDE 8 5 1 GROUND THE HV POWER SUPPLY 8 5 2 SUPPLY 48V POWER 8 5 3 SET REQUIRED OUTPUT HV 8 5 4 CONNECT THE LOAD 8 5 5 CHECK THE DC POWER SUPPLY 9 CHAPTER 6 CONTROL VIA CAN RS232 10 6 1 CAN 10 6 2 RS232 10 CHAPTER 7 CAN BROWSER CONTROLS 11 ...

Page 4: ...RE 3 REAR CONTROLS OF THE HV POWER SUPPLY 5 FIGURE 4 RS232 CAN CONNECTOR PINOUT 6 FIGURE 5 ON OFF CONTROL INPUT 8 FIGURE 6 CAN USB CONVERTER 10 LIST OF TABLES TABLE 1 MODELS 2 TABLE 2 MAIN COMPONENTS 2 TABLE 3 TECHNICAL SPECIFICATIONS 3 TABLE 4 CONTROLS AND CONNECTIONS OF THE HV POWER SUPPLY 5 TABLE 5 BASIC CONTROLS 11 TABLE 6 ADVANCED CONTROL REGISTERS 11 ...

Page 5: ...s into validity since the shipment date Any evidence of improper use or unauthorized repair attempts voids the warranty For service warranty requests please contact EKSMA OPTICS UAB c o EKSMA Optics UAB Dvarcioniu st 2B LT 10233 Vilnius Lithuania Phone Fax E mail Website 370 5 272 99 00 370 5 272 92 99 info eksmaoptics com www eksmaoptics com ...

Page 6: ... 0 45 HV 400 1 8 1 8 220 HV 400 2 6 2 6 152 HV 400 3 6 3 6 110 HV 400 4 0 4 0 90 HV2 200 1 5 1 5 133 HV2 200 2 0 2 0 100 HV2 200 2 6 2 6 76 HV2 200 3 6 3 6 55 Table 2 Main components Component Quantity Notes High voltage HV power supply 1 Connector for DC input 1 24V cable to Pockels cell driver 1 Optionally if delivered in set with Pockels cell driver RS232 cable 1 Optionally if RS232 control is ...

Page 7: ...V output power at maximal output voltage W 200 400 2 200 Supply voltage DC requirements 47 5 55 V 4 5 A 47 5 55 V 9 A Output voltage ripple 0 3 Output voltage control CAN RS232 Internal potentiometer Maximum 24V output current A 1 2 Maximum ambient temperature for operation C 45 Dimensions mm See CHAPTER 3 DEVICE LAYOUT 200 119 76 Weight g 1120 ...

Page 8: ...4 www eksmaoptics com Chapter 3 DEVICE LAYOUT Figure 1 Outline drawing and dimensions of the HV power supply Figure 2 Front controls of the HV power supply ...

Page 9: ...mode 2 LED HV output on 3 LED 48V 4 24V auxiliary output connector 5 1 2 pins for 48V input 5 2 2 pins for GND input 48V 6 1 Test point for HV current monitoring 1V 100mA 6 2 Test point for HV current monitoring 1V 100mA 6 3 Test point for HV output monitoring 1V 1kV 6 4 Test point GND 7 CAN RS232 connector 8 Grounding screw 9 HV output 10 GND output 11 HV output if equipped ...

Page 10: ...6 www eksmaoptics com Figure 4 RS232 CAN connector pinout ...

Page 11: ... dry non conductive POLLUTTION occurs Warning The safety of the system incorporating driver and HV power supply is the responsibility of the assembler of the system Operating the power supply is allowed to persons acquainted with the operation manual and having permission to deal with voltages over 1000 V Do not remove unit covers while power cable is connected to the mains if applicable WARNING D...

Page 12: ...uned by 1 Using internal 1 trimmer potentiometer Output voltage and current control is available on test point TP OUT connector pins 6 2 Using CAN or RS232 see CHAPTER 6 CONTROL VIA CAN RS232 for details 3 HV output can be switched off by connecting pin 1 of connector CAN RS232 7 to GND pins 5 6 Figure 5 ON OFF control input Switch the DC power supply off Connect the load Check that the connection...

Page 13: ...tch the 48 VDC power on Measure the voltage on 48 VDC input of HV power supply while HV power supply is loaded If the voltage measured is below 47V use thicker wires or increase the DC supply voltage to compensate voltage drop ...

Page 14: ...r supply by cables supplied 2 Turn on 48V power supply 3 Install and open CAN Browser 4 Set required settings see CHAPTER 7 CAN BROWSER 1 Connect HV power supply to PC by the RS232 cable supplied Note do not attempt using ordinary RS232 cables Some of them are not compatible with this series of HV power supplies 2 Turn on 48V power supply 3 Install and open CAN Browser 4 Set required settings see ...

Page 15: ...V supply Tpcb Display PCB temperature of HV supply Vpwr Display input 48 voltage Vprog Display maximal HV value of output for potentiometer mode Interlock Vpwr max Display actual value of 48V maximal input If DC input is over the maximal value see Table 3 HV power supply turns off Interlock Uout max Display setting of maximal HV output value If this value is exceeded HV power supply is off Interlo...