DK50
DK50
PLUS
NP-DK50_5
- 28 -
10/2009
Dental compressor with base DK50 Z, DK50-10 Z, DK50-10 Z/M, DK50 PLUS, DK50 PLUS/M
(Fig.8)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Direct the output pressure hose, drain hose and power cord out the back of the compressor.
Dental compressor in box DK50 S, DK50-10 S
(Fig.5,Fig.8)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Direct the output pressure hose, drain hose and power cord out the back of the compressor. Slide
the box over the compressor so that the front face of the box matches the front part of the compressor and
the box is fully seated. Make sure that the pressure hose, drain hose and electric cord come out via the
opening at the back of the box. Position the drain hose with its valve in the holder at the rear of the box.
Dental compressor in box
DK50-10 S/M
(Fig.5, Fig.8)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Direct the output pressure hose, drain hose and power cord out the back of the compressor. Fit
the housing over the top of the compressor, connect the flexible shaft to the control button, fasten with the
screw and put the lid on the cabinet housing (see picture). Make sure that the pressure hose, drain hose and
electric cord come out via the opening at the back of the box. Position the drain hose with its valve in the
holder at the rear of the box.
Dental compressor in box DK50 PLUS S
(Fig.6, Fig.8)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Place the wall-mounted stopper (28) - 2 pcs onto the compressor housing in the rear top part of
the housing and put the housing onto a required place. The stoppers provide a sufficient distance of the
housing from a wall for thorough ventilation. For setting up the compressor in housing you must open the
door on the housing using the attached key and remove connecting reinforcement (27) in the front bottom
part of housing. If necessary, the door may be disassembled using door hinge (32). Protrude pressure hose
via a hole in housing and connect it to the appliance in a suitable way. Grasp the compressor at its handle
and put it into the casing using built-in wheels (33). Fix the hose of a manometer in a hose into the fast-on
coupling on a compressor, put the connecting reinforcement (27) back and connect the pressure hose to a
compressor. Insert the electric power cord of a compressor into a socket (34) on a housing. By slight rotation
of rectification screws (44) set the correct position of door against the casing frame. When closing the door
the pin (45) on the door must easily snap in the opening in the casing frame. Close the housing doors and
duly lock the lock (26). Connect the mains plug into the mains socket.
It is not allowed to leave the key in a lock! It must be saved against non-instructed persons!
Dental compressor in box DK50 PLUS S/M
(Fig.6, Fig.8)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Place the compressor into housing similarly as in the previous paragraph. Prior to placing the
compressor into housing, protrude house for condensate drain via hole in housing and connect it to a bottle
(22). The magnetic holder (31) with a vessel (22), for entrapping condensate from a dryer may be fixed at the
sides of housing or from the front on its doors. When fixing the holder with a vessel at the housing side it is
necessary to consider a space of at least 11 cm between the housing and furniture. Distance smaller than
the specified one may cause problem with handling of the vessel.
10.2. Compressed air outlet
(Fig.9)
At the compressed air outlet (1) of the compressor, connect the pressure
hose with a nut (2) and fasten clip (3). Connect the hose to the dental
appliance.
10.3. Electrical connection
Plug the electrical cord into the mains.
The appliance is equipped with a grounded plug. Make sure this connection complies with
local electrical codes. The mains voltage and frequency must comply with the data stated
on the appliance label.
•
Keep the socket easily accessible to ensure that in an emergency the appliance can be safely
disconnected from the mains.
•
Connection to the power distribution box must be max.16 A.
Electrical cable may not contact the hot parts of a compressor. Insulation could be
damaged!
If any electrical cord or air hose is damaged it must be replaced immediately.
Fig.9
Summary of Contents for DK50 Plus Mobile
Page 1: ......
Page 2: ......
Page 37: ...DK50 DK50 PLUS 10 2009 35 NP DK50_5 1 CE CE 93 42 EEC 2 2 1 2 2...
Page 38: ...DK50 DK50 PLUS NP DK50_5 36 10 2009 2 3 3 CE...
Page 44: ...DK50 DK50 PLUS NP DK50_5 42 10 2009 3 MONZUN M1a 4 MONZUN M1 10 5...
Page 45: ...DK50 DK50 PLUS 10 2009 43 NP DK50_5 6 DK50 PLUS S 7 KJF1...
Page 46: ...DK50 DK50 PLUS NP DK50_5 44 10 2009 8 X Y 9 5 C 40 C 70 6 10 10 1...
Page 52: ...DK50 DK50 PLUS NP DK50_5 50 10 2009 16 10 1 11 12 3 2 10 17 1 11 12 18 19...
Page 104: ......