4.
Over all the supply has four outputs, out of which three are the DC outputs & one is
the Low Voltage AC 6.0V, 2 Amps.
Technical Data :
Fuse Rating:
F 3.15A / 250V
Mains Input:
Selectable120 (50/60Hz) or 240 (50/60Hz)
Max. Output Wattage:
140 Watt Max.
DC :
Voltage:
0-20V (Regulated), 5 Amps. Max. adjustable from 0 to +20V the
output voltage is displayed on the DPM, keeping A/V selector
switch in the V (Voltage Position). To know the current, set the
selector switch to A ( Amps.) position.
10 - 500V, 0.01 Amp. Max. adjustable from 10 to +500V. The output
voltage is displayed on the DPM keeping the selector switch
towards 500V markings.
0-80V, 0.01 Amp. Max. adjustable from 0 to +80V. The output
voltages is displayed on the DPM keeping the selector switch
towards the 80V markings.
Important Note :
While using H-Voltages 10-500V & 0-80V, please note down the following instruction
very Carefully.
1.
Always turn the voltage controls knobs (Grey for 10-500V & Yellow for 0-80V)
fully anti Clockwise to make the O/P voltages 000V.
2.
Connect the required Load / Under Test apparatus at the appropriate O/P
sockets.
3.
Then slowly turn the voltage control knobs in the clockwise direction to set the
desired O/P voltage which is also displayed on the Digital Voltmeter.
4.
Before removing the load / apparatus from the O/P sockets first of all alwasys
make the O/P Voltage 000, only then remove the connections from O/P sockets,
for the safe guard of DPM / equipment to avoid any electric shock due to HV DC
available at the O/P Sockets.
AC :
Voltages
6V (Max.) selectable from 1 to 6V as per the markings.
Current
2 Amps. Max.
Caution :
1.
Replace the fuse as per the specified rating & type.
2.
Do not over-load more than 5 Amps.
-2-
© EISCO SCIENTIFIC
●
www.eiscolabs.com
!