REPLACEMENT OF FAULTY PARTS/COMPONENTS
Metal Heater (Part No. 4) Replacement:
1.
The Metal Heater (Part No. 4) is turned off and the still is
disconnected from the mains electric supply.
2.
The Glass Boiler (Part No. 2) is cleaned if heavy scaling is present by
the following cleaning procedures detailed in this manual.
3.
It is easier to replace the heater by working with the Glass Boiler on
the bench
*
Remove the Glass Condenser (Part No. 3)
*
Drain the Glass Boiler (Part No. 2) via drain cock
*
Remove the Glass Boiler (Part No. 2) complete with Metal Heater
(Part No. 4) from the S.S. stand (Part No. 1)
*
Unscrew the three stainless steel bolts (Part No. 9) securing the
Metal Heater (Part No. 4) to the Glass Boiler (Part No. 2)
*
The new Metal Heater is fitted to the Glass Boiler by following the
installation instructions - 3
4.
The still is reassembled by following the installation instructions 4 -
12.
CIRCUIT DIAGRAM
-9-
© EISCO SCIENTIFIC
●
www.eiscolabs.com
GREEN
BLACK
RED