For maximum screening accuracy, temperature
calibration should be done every time the device is
turned on, or when the device is moved to a different
location with different ambiance temperature.
Please operate using individuals with normal body
temperature (+- 36 deg C). If possible, use a normal
thermometer to measure the body temperature first.
Once normal body temperature is confirmed, follow
these steps to complete the temperature calibration:
1. Press and hold anywhere on the screen for a few
seconds.
2. On the device password entry screen, click
“
Temperature Calibration
”.
* Please complete selection before the
countdown reach 0.
3. Select “
Auto
” for automatic calibration or
select “
Manual
” for precise calibration
(recommended).
4. Click “
Start Calibration
”.
A. TEMPERATURE CALIBRATION