background image

11      EMC  Tables

 

 

                                                                                                    Page    23  /  25 

Recommended separation distances between portable and mobile RF communications 
equipment and the Fast-Pack. 

The 

Fast-Pack 

is intended for use in an electromagnetic environment in which radiated RF 

disturbances are controlled. The customer or the user of the 

Fast-Pack 

can help prevent 

electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the 

Fast-Pack 

as recommended 

below, according to the maximum output power of the communications equipment.

 

Rated maximum 

output power of 

transmitter 

W

 

Separation distance according to frequency of transmitter 

m

 

150 kHz to 80 MHz 

1.2

d

P

=

 

80 MHz to 800 MHz 

1.2

d

P

=

 

800 MHz to 2.5 GHz 

2.3

d

P

=

 

0.01 

0.12 

0.12 

0.23 

0.1 

0.38 

0.38 

0.73 

1.2 

1.2 

2.3 

10 

3.8 

3.8 

7.3 

100 

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance 

in meters (m) can be estimated using the equation applicable to the 

frequency of the transmitter, where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range 
applies. 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.

 

 

Summary of Contents for Fast-Pack

Page 1: ...Fast Pack USER MANUAL...

Page 2: ......

Page 3: ...ntended Use 7 3 2 Contraindications 7 4 Installing the Fast Pack 8 4 1 Installation of heating needle 8 4 2 Installation of adapter 8 4 3 Connecting charge base 9 5 Use Interface 10 6 Setting 11 6 1 M...

Page 4: ...1 Scope of Fast Pack Page 4 25 1 Scope of Fast Pack 1 1 Parts Identification 1 2 3 4 1 Charge Base 2 Handpiece 3 Heating Needle 3PCS 4 Adapter...

Page 5: ...ase 1pcs Part No 6351003 Adapter 1pcs Part No 6316001 6316005 Heating needle S 1pcs Size 40 0 025 Color Black Part No 6351006 Heating needle M 1pcs Size 50 0 05 Color Yellow Part No 6351007 Heating ne...

Page 6: ...on and performance Serial number Catalogue number Manufacturer Date of manufacture Safety class II device Type B applied part CE marking Direct current Do not dispose of with normal household waste St...

Page 7: ...flammable anesthetic gas mixtures or flammable substances 3 The device must not be placed in humid surroundings or anywhere where it can come into contact with any type of liquids 4 Thermal hazard ris...

Page 8: ...one of 6 orientations Pull it out from handpiece then can be installed in other orientations WARNING To remove the heating needle wait for it to cool down it takes about 2 3 seconds and real time tem...

Page 9: ...the original adapter could be used Put the handpiece all the way into the charge base the charge state will show on the screen NOTE Put the handpiece into the charge base in the right direction otherw...

Page 10: ...Parameter Setting During standby state holding down press then press in 2 seconds to entry memory parameter setting parameter of T1 to T5 can be set independently Heating Long press Turn Power Off Lo...

Page 11: ...e Temperature of this memory can be change Press till target temperature the temperature can be set from 90 to 250 press to confirm Keep heat Time Sec 10 Press again the Keep heat Time of this memory...

Page 12: ...15 minutes press to confirm Beep Volume Vol 1 Press again the Beep Volume can be change press to adjust the Beep Volume can be set 0 1 and 2 press to confirm RestoreSettings NO YES Press again the Re...

Page 13: ...ge base is recommended See chapter 4 3 Connecting charge base NOTE Only the original adapter could be used Charging indication appears on the screen and flashes slowly when battery is fully charged or...

Page 14: ...indicator LED lights up during heating Only the end of heating needle about 4 5mm will be heated use this area to cut the gutta percha 1 2 3 166 2 1 The Keep heat Time will display on screen When the...

Page 15: ...cooling 1 the cooling indication will appear 2 Real time temperature of cooling heating needle will appear When the set time of CoolingDisplay is reached will change to standby interface Tip error If...

Page 16: ...ion pouches Sterilization Steam sterilization at 134 C at least 4 minutes or Steam sterilization at 121 C at least 35 minutes Minimum drying time after sterilization 10 minutes Storage Keep the compon...

Page 17: ...e the components with a piece of gauze that has been dampened with Ethanol for Disinfection Ethanol 70 to 80 vol and wrung out thoroughly NOTE Do not use anything except Ethanol for Disinfection Ethan...

Page 18: ...s not connected Check the connection 4 3 The plug of the adapter is not inserted into the outlet Check the connection There is no electricity in the outlet Check the connection No charge indicator fla...

Page 19: ...2600mAh 10 Charger power supply AC 100 240 V 10 Charger power output 6V 3A Frequency 50 60Hz 10 Temperature 90 250 Electrical safety class Class II Applied part B Operating conditions Use in enclosed...

Page 20: ...including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Not app...

Page 21: ...les 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Not applicable Not applicable Not applicable The test is not applicable since the EUT does not have AC power port Power...

Page 22: ...ng symbol IEC 61000 4 6 V m Radiated RF 3 5 V m IEC 61000 4 3 NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic pr...

Page 23: ...r W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 1 2 d P 80 MHz to 800 MHz 1 2 d P 800 MHz to 2 5 GHz 2 3 d P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3...

Page 24: ...duct are reserved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to CHANGZHOU SIFARY MEDICAL TECHNOLOGY CO LTD The industrial des...

Page 25: ...trict Changzhou City 213000 Jiangsu China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www eighteeth com Landlink GmbH Tel 0049 7641 9626855 Fax 0049 7641 9626855 Add Dorfstrasse...

Reviews: