background image

7      Operation

  

                                                                                                    Page    13  /  25 

7.Operation 

7.1 Charge 

 

Displays  the  present  remaining  amount  of  the 
battery. 

Less than 15% remains, please charge. 

 

NOTE

 

If the power if less than 15%, must be recharged 
within 30 days, otherwise the battery will be 
damaged. 

Alternative charging m ethod

 

Charge without  charge base also available, using 
adapter connect to handpiece directly,  the charge 
state will show on the screen. 

Charge  with  charge  base  is  recommended 

(See 

chapter 4.3 Connecting charge base

错误

!

未找到

引用源。

)

NOTE

 

Only the original adapter could be used. 

           

 

Charging  indication  appears  on  the  screen,  and 
flashes slowly, when battery is fully charged or in a 
state  near  full  charge,  the  flash  will  stop.  Fully 
charged  will  take  about  4  hours,  depending  on 
residual battery power and battery state. 

It can be recharged 300-500 times, depending on 
the operating conditions of the device. 

WARNING

 

Do not change the battery, only trained technician 
or distributor can change the battery, the electronic 
parts  will  be  damaged  if  use  a  wrong  battery  or 
install with a wrong way. 

 

 

 

 

Summary of Contents for Fast-Pack

Page 1: ...Fast Pack USER MANUAL...

Page 2: ......

Page 3: ...ntended Use 7 3 2 Contraindications 7 4 Installing the Fast Pack 8 4 1 Installation of heating needle 8 4 2 Installation of adapter 8 4 3 Connecting charge base 9 5 Use Interface 10 6 Setting 11 6 1 M...

Page 4: ...1 Scope of Fast Pack Page 4 25 1 Scope of Fast Pack 1 1 Parts Identification 1 2 3 4 1 Charge Base 2 Handpiece 3 Heating Needle 3PCS 4 Adapter...

Page 5: ...ase 1pcs Part No 6351003 Adapter 1pcs Part No 6316001 6316005 Heating needle S 1pcs Size 40 0 025 Color Black Part No 6351006 Heating needle M 1pcs Size 50 0 05 Color Yellow Part No 6351007 Heating ne...

Page 6: ...on and performance Serial number Catalogue number Manufacturer Date of manufacture Safety class II device Type B applied part CE marking Direct current Do not dispose of with normal household waste St...

Page 7: ...flammable anesthetic gas mixtures or flammable substances 3 The device must not be placed in humid surroundings or anywhere where it can come into contact with any type of liquids 4 Thermal hazard ris...

Page 8: ...one of 6 orientations Pull it out from handpiece then can be installed in other orientations WARNING To remove the heating needle wait for it to cool down it takes about 2 3 seconds and real time tem...

Page 9: ...the original adapter could be used Put the handpiece all the way into the charge base the charge state will show on the screen NOTE Put the handpiece into the charge base in the right direction otherw...

Page 10: ...Parameter Setting During standby state holding down press then press in 2 seconds to entry memory parameter setting parameter of T1 to T5 can be set independently Heating Long press Turn Power Off Lo...

Page 11: ...e Temperature of this memory can be change Press till target temperature the temperature can be set from 90 to 250 press to confirm Keep heat Time Sec 10 Press again the Keep heat Time of this memory...

Page 12: ...15 minutes press to confirm Beep Volume Vol 1 Press again the Beep Volume can be change press to adjust the Beep Volume can be set 0 1 and 2 press to confirm RestoreSettings NO YES Press again the Re...

Page 13: ...ge base is recommended See chapter 4 3 Connecting charge base NOTE Only the original adapter could be used Charging indication appears on the screen and flashes slowly when battery is fully charged or...

Page 14: ...indicator LED lights up during heating Only the end of heating needle about 4 5mm will be heated use this area to cut the gutta percha 1 2 3 166 2 1 The Keep heat Time will display on screen When the...

Page 15: ...cooling 1 the cooling indication will appear 2 Real time temperature of cooling heating needle will appear When the set time of CoolingDisplay is reached will change to standby interface Tip error If...

Page 16: ...ion pouches Sterilization Steam sterilization at 134 C at least 4 minutes or Steam sterilization at 121 C at least 35 minutes Minimum drying time after sterilization 10 minutes Storage Keep the compon...

Page 17: ...e the components with a piece of gauze that has been dampened with Ethanol for Disinfection Ethanol 70 to 80 vol and wrung out thoroughly NOTE Do not use anything except Ethanol for Disinfection Ethan...

Page 18: ...s not connected Check the connection 4 3 The plug of the adapter is not inserted into the outlet Check the connection There is no electricity in the outlet Check the connection No charge indicator fla...

Page 19: ...2600mAh 10 Charger power supply AC 100 240 V 10 Charger power output 6V 3A Frequency 50 60Hz 10 Temperature 90 250 Electrical safety class Class II Applied part B Operating conditions Use in enclosed...

Page 20: ...including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC61000 3 2 Not app...

Page 21: ...les 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Not applicable Not applicable Not applicable The test is not applicable since the EUT does not have AC power port Power...

Page 22: ...ng symbol IEC 61000 4 6 V m Radiated RF 3 5 V m IEC 61000 4 3 NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic pr...

Page 23: ...r W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 1 2 d P 80 MHz to 800 MHz 1 2 d P 800 MHz to 2 5 GHz 2 3 d P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3...

Page 24: ...duct are reserved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to CHANGZHOU SIFARY MEDICAL TECHNOLOGY CO LTD The industrial des...

Page 25: ...trict Changzhou City 213000 Jiangsu China Tel 86 0519 85962691 Fax 86 0519 85962691 Email ivy sifary com Web www eighteeth com Landlink GmbH Tel 0049 7641 9626855 Fax 0049 7641 9626855 Add Dorfstrasse...