Maintenance
Page 35 / 52
·
5 min intermediate rinsing with warm water
(>40°C);
·
emptying
The automated cleaning processes have been
validated by using 0.5% neodisher MediClean
forte (Dr. Weigert).
Note Acc. to EN ISO 17664 no manual
reprocessing methods are required for these
devices. If a manual reprocessing method has to
be used, please validate it prior to use.
WARNING
1. Use only approved washer-disinfectors
according to EN ISO 15883, maintain and
calibrate it regularly.
2. Follow instructions and observe
concentrations given by the manufacturer (see
general recommendations).
3. Avoid any contact between the contra angle
and any instrument, kit, support or container.
Disinfection
Automated Thermal Disinfection in
washer/disinfector under consideration of
national requirements in regards to A0 value (see
EN ISO 15883).
A disinfection cycle of 5 min disinfection at 93°C
has been validated for the device to achieve an
A0 value of 3000.
After manual cleaning, the instruments should be
automated disinfected of sterilized immediately.
A manual disinfection is not recommended.
Drying
Automated Drying:
Drying of outside of instrument through drying
cycle of washer/disinfector. If needed, additional
manual drying can be performed through lint
free towel. Insufflate cavities of instruments by