12
8.
Function check
Leakage test
Connect the leakage tester to videoscope and rotate tightly. Screw down the deflated knob and
press the ball. Immerse the videoscope in the asepsis solution, keep the pressure 22 KPa for 30
seconds and observe the bubbles in solution. If there is no bubble and the finger of pressure
meter is changed within 2, it is seal. If it is more than 2, it is leakage somewhere. Do not use
the videoscope again and contact service. After leakage testing, take out the videoscope from
the solution. Screw off the deflated knob to the finger of presser meter to zero. Take off the
leakage tester from videoscpe. It would be damaged without deflation.
The adaptor of leakage tester The deflated knob
Note
- Before each operation please do leakage testing. Please do not use the videoscope if there
is any leakage.
- After each leakage testing, take out the videoscope from the solution, screw off the deflated
knob. After deflating, take off the leakage testing. Otherwise the videoscope would be
damaged.