Cleaning and Maintenance of Anesthaesia Equipment
Remove breathing bag, the patient tube and other tubes from the device and dry them (if necessary, clean with
detergent).
Unscrew the lids of the breathing valves and let them dry too.
If you remove the valve plates, it will speed up the drying process. But please make sure to insert these plates again
before the next anaesthetic, otherwise proper gas exchange is no longer guaranteed.
Approximately every 10 hours, please change the absorbent, even if the indicator change did not take place.
Please check every 20 hours of anesthaesia all exposed sealing rings for wear and tear and lubricate with silicon
vaseline.
Once a year, the evaporator should be subjected to a technical inspection to ensure proper functioning. In this case,
please contact the EICKEMEYER® Customer Service.
Precautions for handling Isoflurane:
Isoflurane is a highly volatile anaesthetic. Therefore, when used in closed spaces ensure that a suction tube is
properly connected to the pressure relief valve or the Bain Circuit. If ventilation of gases is not possible into the
open air, we recommend an anesthetic gas filter (Art. No. 213485).