Eickemeyer 213062 Manual Download Page 8

Manual 

EICKEMEYER

®

 NarkoVet                                                                            

 

6/9 

5

 

Preparing the Device before Anesthaesia 

Check the oxygen supply (pressure and access to sufficient oxygen).

 

 
Connect patient tube and breathing bag to the anesthesia machine and close the pressure relief valve. 
Keep your thumb on the patient's end of the tube and inflate the breathing bag via the  
Oxygen bypass until the breathing bag is bulging. The anesthaetic system is sufficiently tight once a fresh gas flow 
of less than 200 ml is needed to maintain this inner pressure.

 

Now check the function of the overpressure valve by fully opening it very slowly. 

 

If the internal pressure is not adequately maintained, there is a leak in the system  which has to be found before 
using the anaesthetic machine.

 

Check the anesthetic level of the Evaporator.

 

 
 

A common leakage location of the evaporator is the filling tunnel with the seal plate. The rubber rings can be brittle 
and  disintegrate  so  that  the  seal  is  no  longer  functioning.  As  well  an  insufficiently  tightened  sealing  block  may 
cause anesthetic loss.

 

In general, due to the built in pressure compensation of the evaporator a loss of anesthetic agent up to 0.5 cm

3

 per 

day can be expected.

 

 

 
 
 

6

 

Anaesthetic Procedure 

Typically  in  a  semi-closed  system  a  fresh  gas  flow  of  2  l/min  at  a  selected 

concentration  of  3.0  vol%

  at  the 

evaporator is sufficient for a quick wash-in. The depth of anesthaesia needs to be monitored and the concentration 
to be changed if necessary.

 

 
 
To maintain the anaesthesia, generally it is sufficient to follow the guidelines of the MAC- 15% inspiratory 
add as per the package leaflet at minimum 500 ml/min oxygen inflow (watch status of anaesthesia and breathing 
continuously). 

 
The MAC values are 

 

Dog:    

 

 

1.28% 

Cat:  

 

 

 

1.63% 

Birds:    

 

 

1.45% 

Small mammals:   1.34% 
Rat:  

 

 

 

2.40%

 

 
Pay  attention  to  the  decreased  wash-in  and  wash-out  times  of  an  Isoflurane-only  anesthaesia  in  comparison  to 
halothane  or  mixed  anesthaesia.  In  general,  the  evaporator should  be  closed  just  at  the  time  of  the  final  suture       
(5 min before the intended end of the OP).

 

 
 

 
 
 
 

Summary of Contents for 213062

Page 1: ...MANUAL TELEPHONE 49 7461 96 580 0 www eickemeyer com NARKOVET ANAESTHETIC UNIT Art No 213062 ...

Page 2: ......

Page 3: ...tents 1 Description 2 2 Overview of the Device 2 3 Startup Procedure 3 4 Operation of the Pendulum System Bain Circuit 5 5 Preparing the Device before Anesthaesia 6 6 Anaesthetic Procedure 6 7 Cleaning and Maintenance of Anesthaesia Equipment 7 ...

Page 4: ...r that is easy to refill and install with a capacity up to 10 hours of anesthesia Pipes for fresh gas that are easy to adapt and which can also be used for anesthaesia boxes Flow meter that is easy to adjust and read with a gas flow from 0 2 to 4 2 l min Flush button for the oxygen bypass 2 Overview of the Device 1 Precision Flow Meter 2 Oxygen Bypass 3 Evaporator 4 Safety Filling Device 5 Oxygen ...

Page 5: ...l the evaporator with Isoflurane The procedure is shown on the evaporator 1 Screw the Isoflurane key fill adapter onto bottle Art No 213246 2 Untighten the screw on the side of the safety filling device remove the sealing block introduce the safety filler into the filling tunnel and tighten screw 3 Turn the black filling screw to the 12 clock position 4 Fill the evaporator up to the maximum fillin...

Page 6: ...tainer 11 counterclockwise Remove the soda lime Fill the container with soda lime The max filling line must not be exceeded Insert the container and lock it by turning the lock lever clockwise Plug in the patient circuit and the breathing bag Connect an exhaust pipe to the pressure relief valve ...

Page 7: ...thing bag and in the case of overpressure air is transferred via the open valve in the exhaust pipe The pressure relief valve should always be fully opened during spontaneous breathing against clockwise direction A re breathing of exhaled carbon dioxide is prevented by the influx of fresh gas into the breathing bag open system Recommended quantity of fresh gas for wash in 800 ml 1000 ml min Concen...

Page 8: ...grate so that the seal is no longer functioning As well an insufficiently tightened sealing block may cause anesthetic loss In general due to the built in pressure compensation of the evaporator a loss of anesthetic agent up to 0 5 cm3 per day can be expected 6 Anaesthetic Procedure Typically in a semi closed system a fresh gas flow of 2 l min at a selected concentration of 3 0 vol at the evaporat...

Page 9: ...rs please change the absorbent even if the indicator change did not take place Please check every 20 hours of anesthaesia all exposed sealing rings for wear and tear and lubricate with silicon vaseline Once a year the evaporator should be subjected to a technical inspection to ensure proper functioning In this case please contact the EICKEMEYER Customer Service Precautions for handling Isoflurane ...

Page 10: ...Manual EICKEMEYER NarkoVet 8 9 Notes ...

Page 11: ...Manual EICKEMEYER NarkoVet 9 9 Notes ...

Page 12: ... www eickemeyer nl DENMARK EICKEMEYER ApS Lysbjergvej 6 Hammelev 6500 Vojens T 45 7020 5019 F 45 7353 5019 E info eickemeyer dk www eickemeyer dk UNITED KINGDOM EICKEMEYER Ltd 47 St Margarets Grove Twickenham Greater London TW1 1 JF T 44 20 8891 2007 F 44 20 8891 2686 E info eickemeyer co uk www eickemeyer co uk ITALY EICKEMEYER S R L Via G Verdi 8 65015 Montesilvano PE T 39 0859 35 4078 F 39 0859...

Reviews: