26
A. DC Voltage
Range
Resolution
Accuracy
Overload Protection
400mV
100µV
±(0.8%+3)
500V DC or AC
4V
1mV
±(0.8%+1)
40V
10mV
400V
100mV
500V
1V
±(1%+3)
Remark:
Input impedance: 10MΩ
B. AC Voltage
Range
Resolution
Accuracy
Overload Protection
4V
1mV
±(1.2%+3)
500V DC or AC
40V
10mV
400V
100mV
500V
1V
±(1.5%+5)
Remarks:
- Input impedance: approx. 10MΩ.
- Displays effective value of sine wave (mean value response).
- Frequency response 40Hz ~ 400Hz.
C. DC Current
Range
Resolution
Accuracy
Overload Protection
400µA
0.1µA
±(1%+2)
500mA/250V fast type
fuse Ø5x20mm
4000µA
1µA
40mA
10µA
±(1.2%+2)
400mA
100µA
4A
1mA
±(1.5%+5)
Un-fused
10A
10mA
Remark:
- At 10A Range: For continuous measurement 10 seconds and interval not less than 15 minutes.
- Input impedance: approx. 10MΩ.
- Measurement voltage drop: Full range at 400mV.