61
E. Resistance
Range
Resolution
Accuracy
Overload Protection
COD. 51253
COD. 51254
200
0.01
±(0.5%+20)+
test leads open
circuit value
±(0.4%+20)+
test leads open
circuit value
1000V
2k
0.0001k
±(0.5%+20)
±(0.4%+20)
20k
0.001k
200k
0.01k
±(1%+20)
±(0.8%+20)
2M
0.0001M
±(1%+40)
±(1%+40)
20M
0.001M
±(1.5%+40)
±(1.5%+40)
F. Continuity Test
Range
Resolution
Overload Protection
0.01
1000V
Remarks:
- Open circuit voltage approximate -1.2V.
- The buzzer does not sound when the test resistance is >60 .
- The beeper comes on continuously for open conditions, that is test resistance is ≤40
.
G. Diode Test
Range
Resolution
Overload Protection
0.0001V
1000V
Remarks:
- Open circuit voltage approximate 2.8V.
- A good silicon junction drops between 0.5V and 0.8V.
H. Capacitance
Range
Resolution
Accuracy
Overload Protection
COD. 51253
COD. 51254
20nF
0.001nF
±(1.5%+20)+
capacitance
value of open
circuit test
leads
±(1.2%+20)+
capacitance
value of open
circuit test
leads
1000V
200nF
0.01nF
±(1.5%+20)
±(1.2%+20)
2µF
0.0001µF
20µF
0.001µF
±(1.5%+40)
±(1.2%+40)
200µF
0.01µF
±(1.5%+40)
2mF
0.0001mF
±(5%+40)
±(5%+40)
20mF
0.001 mF
Not specified
Not specified