28
- When the measuring current >1A, press
Hz
button to toggle between AC current and
frequency measurement mode. But the frequency readings obtained from this range is only for
reference.
- Change to AC:
Couple AC and True RMS response method. Input sine wave to adjust. Non sine wave must
follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy.
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
SPECIFICATIONS
A. General Specifications:
-
Display
: 3 3/4 digits LCD display, Maximum display 3999.
-
Polarity
: Auto
-
Overloading
: Display
OL
or –
OL
.
-
Battery Deficiency
: Display
.
-
Sampling
: 3 times per second.
-
Measurement Deviation
: The conductor being meaured is not placed in the center of the jaw
during AC/DC current measurement, it will cause extra ±1% deviation based on the stated
accuracy.
-
Drop Test
: 1 meter drop test passed.
-
Max. Jaw Size
: 55mm diameter.
-
Projected Max. Current conductor size
: 45mm diameter.
-
Electro-Magnetic
: When carrying out measurement near the electromagnetic, it may cause
unstable or wrong reading.
-
Power
: 1 x 9V battery (6LF22 1604A)
-
Dimensions
: 285.3mm x 105mm x 44.5mm
-
Weight
: Approximate 533g (battery included)
B. Environmental Requirements
- The Meter is suitable for indoor use.
-
Altitude
: Operating: 2000m
Storage: 10000m
-
Safety/ Compliances
: IEC 61010 CATII 600V, CATIII 300V over voltage and
double insulation standard, pollution degree 2.
-
Temperature and humidity
:
Operating:
0ºC~30ºC(≤85%R.H)
30ºC~40ºC (≤75%R.H)
40ºC~50ºC(≤45%R.H)
Storage:
-20ºC~+60ºC (≤85%R.H)